FDA Adverse Event Malfunction Summary report: N

2243471-2019-00013

MDR report key: 8440123 · Received March 21, 2019

Report

Report Number
2243471-2019-00013
Event Type
Malfunction
Date Received
March 21, 2019
Date of Event
January 26, 2019
Report Date
March 21, 2019
PMA / PMN Number
BK140195
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

FOR THE CUSTOMER'S INSTRUMENT (SERIAL ID SN-(B)(4)), ALL STOP DISCS AND O-RINGS WERE REPLACED ON PROCESSING MODULE B ON THE 15 JANUARY 2019, AND ARE PLANNED TO BE REPLACED ON THE PROCESSING MODULE A. THE MATERIAL HAS BEEN REQUESTED BACK FOR FURTHER INVESTIGATION WITH THE SUPPLIER. THE DROPLETS OBSERVED AROUND THE HEATING STATIONS WERE CLEANED BY THE LOCAL FIELD SERVICE ENGINEER. CORRECTIVE AND PREVENTIVE ACTIONS WILL BE IMPLEMENTED AS APPROPRIATE. (B)(4).

Description of Event or Problem · 1

A (B)(6) FIELD SERVICE ENGINEER (FSE) PERFORMED A TIGHTNESS CHECK ON A CUSTOMER'S COBAS 8800 SYSTEM (SERIAL ID (B)(4)), WHICH FAILED FOR PROCESSING MODULE A, ON (B)(6) 2019. SUBSEQUENT CHECKS ON THIS SYSTEM PASSED ON (B)(6) 2019. REVIEW OF CUSTOMER-PROVIDED DATA SHOWED ONE (1) P07P ERROR CODE WAS GENERATED FOR A SAMPLE PROCESSED ON THE COBAS 8800 SYSTEM PROCESSING MODULE A ON (B)(6) 2019; THE ERROR CODE P07P IS INDICATIVE OF A VOLUME ERROR DURING SUPERNATANT REMOVAL. WHEN ERROR CODES ARE GENERATED FOR SAMPLES, THE SAMPLES ARE INVALIDATED AND NEED TO BE REPEAT TESTED PER THE INSTRUCTIONS FOR USE. ADDITIONALLY, IMAGES ON THE PROCESSING MODULE DECKS WERE PROVIDED AND SHOWED EVIDENCE OF LEAKAGE/DROPLETS ON THE FRONT HEATING STATION. NO HARM OR INJURY WAS ALLEGED WITHIN THE CASE.

Patients

Seq Age Sex Outcome Treatment
1