2243471-2019-00013
Report
- Report Number
- 2243471-2019-00013
- Event Type
- Malfunction
- Date Received
- March 21, 2019
- Date of Event
- January 26, 2019
- Report Date
- March 21, 2019
- PMA / PMN Number
- BK140195
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
FOR THE CUSTOMER'S INSTRUMENT (SERIAL ID SN-(B)(4)), ALL STOP DISCS AND O-RINGS WERE REPLACED ON PROCESSING MODULE B ON THE 15 JANUARY 2019, AND ARE PLANNED TO BE REPLACED ON THE PROCESSING MODULE A. THE MATERIAL HAS BEEN REQUESTED BACK FOR FURTHER INVESTIGATION WITH THE SUPPLIER. THE DROPLETS OBSERVED AROUND THE HEATING STATIONS WERE CLEANED BY THE LOCAL FIELD SERVICE ENGINEER. CORRECTIVE AND PREVENTIVE ACTIONS WILL BE IMPLEMENTED AS APPROPRIATE. (B)(4).
A (B)(6) FIELD SERVICE ENGINEER (FSE) PERFORMED A TIGHTNESS CHECK ON A CUSTOMER'S COBAS 8800 SYSTEM (SERIAL ID (B)(4)), WHICH FAILED FOR PROCESSING MODULE A, ON (B)(6) 2019. SUBSEQUENT CHECKS ON THIS SYSTEM PASSED ON (B)(6) 2019. REVIEW OF CUSTOMER-PROVIDED DATA SHOWED ONE (1) P07P ERROR CODE WAS GENERATED FOR A SAMPLE PROCESSED ON THE COBAS 8800 SYSTEM PROCESSING MODULE A ON (B)(6) 2019; THE ERROR CODE P07P IS INDICATIVE OF A VOLUME ERROR DURING SUPERNATANT REMOVAL. WHEN ERROR CODES ARE GENERATED FOR SAMPLES, THE SAMPLES ARE INVALIDATED AND NEED TO BE REPEAT TESTED PER THE INSTRUCTIONS FOR USE. ADDITIONALLY, IMAGES ON THE PROCESSING MODULE DECKS WERE PROVIDED AND SHOWED EVIDENCE OF LEAKAGE/DROPLETS ON THE FRONT HEATING STATION. NO HARM OR INJURY WAS ALLEGED WITHIN THE CASE.
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |