FDA Adverse Event Malfunction Summary report: N

HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT

MDR report key: 8439895 · Received March 21, 2019

Report

Report Number
3005180920-2019-00171
Event Type
Malfunction
Date Received
March 21, 2019
Date of Event
February 19, 2019
Report Date
March 21, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
UDI-DI
07630040742115
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 18 MARCH 2019: LOT 1854515: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-DEC-2018. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ANOTHER SIMILAR EVENT HAS BEEN REPORTED ON THIS LOT (COMPLAINT (B)(4)). VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER: THE SPRING PLUNGER IS BROKEN. IT IS POSSIBLE THAT THIS OCCURRENCE COULD HAVE BEEN INFLUENCED BY BAD ASSEMBLY DURING PRODUCTION, HOWEVER THIS CANNOT BE CONFIRMED.

Description of Event or Problem · 1

DURING THE SURGERY THE BROACH HANDLE BALL SPRING FELL OUT OF ITS SIEGE. IT DID NOT FALL IN THE PATIENT. THE SURGEON USED THE STRAIGHT BROACH HANDLE TO COMPLETE THE SURGERY SUCCESSFULLY, NO CRITICAL DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234740 HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT SURGICAL INSTRUMENT FOR HIP SURGERY LXH MEDACTA INTERNATIONAL SA 1854515 07630040742115

Patients

Seq Age Sex Outcome Treatment
1 Other