HEMOCUE HB 201+ SYSTEM
Report
- Report Number
- 3003044483-2019-00006
- Event Type
- Malfunction
- Date Received
- March 21, 2019
- Date of Event
- January 4, 2019
- Report Date
- March 21, 2019
- Manufacturer
- HEMOCUE AB
- Product Code
- GKR
- PMA / PMN Number
- K032203
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
THE ANALYZER WAS SENT TO HEMOCUE AB WHERE THE CUSTOMER PROBLEM WAS CONFIRMED, IT WAS DISCOVERED THAT THE EEPROM MEMORY HAD BEEN CORRUPTED. THE ROOT CAUSE WAS FOUND TO BE A GLITCH IN THE CONNECTION BETWEEN COMPONENTS ON THE MAIN BOARD WHICH HAS CAUSED READ AND WRITE ERRORS LEADING TO CORRUPT VALUES IN EEPROM MEMORY. THE ERROR IN EEPROM MEMORY WILL BE PERMANENT IF THE ROOT CAUSE IS HAPPENING IN COMBINATION WITH OTHER IDENTIFIED SCENARIOS LEADING TO POTENTIAL INCORRECT MEASUREMENTS VALUES. CORRECTIVE ACTION IS TO IMPROVE THE SOLDERING PROCESS OF COMPONENTS ON THE MAIN BOARD. NOTE, WHEN THE ROOT CAUSE IS NOT OCCURRING IN A COMBINATION WITH THE IDENTIFIED SCENARIOS IT WILL LEAD TO AN ERROR CODE AND NO RESULTS WILL BE DISPLAYED.
HEMOCUE AB RECEIVED A COMPLAINT THAT THE ANALYZER WAS SHOWING ERROR CODE E05.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 234270 | HEMOCUE HB 201+ SYSTEM | HEMOGLOBIN TEST SYSTEM | GKR | HEMOCUE AB | 121769 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |