FDA Adverse Event
Malfunction
Summary report: N
PLATO BPS
MDR report key: 84380
·
Received April 14, 1997
Report
- Report Number
- 1121753-1997-00006
- Event Type
- Malfunction
- Date Received
- April 14, 1997
- Date of Event
- December 4, 1996
- Report Date
- April 9, 1997
- Manufacturer
- NUCLETRON CORP.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
AN INCORRECT CALCULATION OF SOURCE DECAY IS POSSIBLE IF, DURING MANUAL PROGRAMMING OF THE MHDR UNIT, THE TREATMENT PLAN IS PRE-PREPARED, THE TREATMENT DATE IS ENTERED AND THIS DATE IS EARLIER THAN THE SOURCE CALIBRATION DATE. IF THESE PARAMETERS ARE ENTERED IN THIS ORDER, THE SOURCE ACTIVITY IS DECAYED RATHER BEING CORRECTED FOR DECAY. THIS DOES NOT OCCUR IF THE PROGRAM CARD IS UTILIZED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLATO BPS | RADIOTHERAPY TREATMENT PLANNING SYSTEM | IWB | NUCLETRON CORP. | PLATO BPS V 13.0 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |