FDA Adverse Event
Malfunction
Summary report: N
PLATO RTS
MDR report key: 84367
·
Received April 14, 1997
Report
- Report Number
- 1121753-1997-00007
- Event Type
- Malfunction
- Date Received
- April 14, 1997
- Date of Event
- March 13, 1997
- Report Date
- April 9, 1997
- Manufacturer
- NUCLETRON CORP.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
ISODOSE DISTRIBUTION IS SHIFTED WHEN THE GRID SIZE IS CHANGED AND ELECTRON BEAMS ARE USED OR WHEN GRID SIZE IS CHANGED AND ELECTRON AND PHOTON BEAMS ARE USED IN COMBINATION. IOSDOSE DISTRIBUTION IS NOT SHIFTER WHEN THE GRID IS CHANGED AND PHOTON BEAMS ALONE ARE USED. THE PROBLEM ONLY OCCURS IS THE FOLLOWING SEQUENCE IS USED: (1) SETUP BEAMS EITHER ELECTRON OR ELECTRON PLUS PHOTON, (2) CALCULATE ALL PANES AND 3) CHANGE THE GRID SIZE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLATO RTS | RADIOTHERAPY TREATMENT PLANNING SYSTEM | IWB | NUCLETRON CORP. | PLATO RTS V. 1.5 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |