STEM: QUADRA-C CEMENTED, MIRROR POLISHING STD STEM SIZE 1 12/14
Report
- Report Number
- 3005180920-2019-00165
- Event Type
- Injury
- Date Received
- March 20, 2019
- Date of Event
- February 19, 2019
- Report Date
- March 20, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030802225
- PMA / PMN Number
- K083558
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR: HIP REVISION SURGERY OCCURRED 4 YEARS AND 8 MONTHS AFTER IMPLANTATION OF A HYBRID DOUBLE-MOBILITY TOTAL HIP ARTHROPLASTY. NO INFORMATION CONCERNING PATIENT AGE, BMI AND GENERAL HEALTH STATUS IS AVAILABLE. BOTH FEMORAL (CEMENTED) AND ACETABULAR (CEMENTLESS) COMPONENTS WERE FOUND TO BE LOOSE, A RATHER UNUSUAL SITUATION. BASED ON THE INFORMATION AVAILABLE, NO FINAL CONCLUSION CAN BE DRAWN. BATCH REVIEW PERFORMED ON 18 MARCH 2019: LOT 111149: 40 ITEMS MANUFACTURED AND RELEASED ON 29-APR-2011. EXPIRATION DATE: 2016-03-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL THE ITEMS OF THIS LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED IN LOOSENING: CUP: VERSAFITCUP 01.26.50MB ACETABULAR SHELL CC Ø 50 (K083116) LOT 135596: 60 ITEMS MANUFACTURED AND RELEASED ON 13-FEB-2014. EXPIRATION DATE: 2019-01-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL THE ITEMS OF THIS LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
ON (B)(6) 2019 WE WERE INFORMED ABOUT A PATIENT WHO CAME IN COMPLAINING OF PAIN 4 YEARS AND 8 MONTHS AFTER THE PRIMARY SURGERY DUE TO AN ASEPTIC LOOSENING OF THE STEM AND A LOOSE CUP. THE CAUSE OF THE LOOSE CUP IS UNKNOWN. ON THE FOLLOWING DAY, THE SURGEON REVISED THE CUP, LINER, HEAD AND STEM AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230872 | STEM: QUADRA-C CEMENTED, MIRROR POLISHING STD STEM SIZE 1 12/14 | CEMENTED HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 111149 | 07630030802225 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |