FDA Adverse Event Malfunction Summary report: N

ACTIVELIFE

MDR report key: 8432224 · Received March 19, 2019

Report

Report Number
9618003-2019-01635
Event Type
Malfunction
Date Received
March 19, 2019
Manufacturer
CONVATEC DOMINICAN REPUBLIC INC
Product Code
EZQ
UDI-DI
00768455181230
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS EMDR IS BEING SUBMITTED TO INCLUDE THE BELOW: D4: UNIQUE IDENTIFIER (UDI). H6: INVESTIGATION RESULTS UNDER IMDRF CAUSE INVESTIGATION CODE, IMDRF INVESTIGATION FINDINGS, IMDRF CAUSE CONCLUSIONS. H10: INVESTIGATION SUMMARY: BATCH RECORD REVIEW LOT 8B00896 WAS MANUFACTURED ON 02/13/2018 IN THE CONVEX 1PC LINE WITH A TOTAL OF (B)(4) MARKET UNITS. COMPLAINT INVESTIGATOR PERFORMED A BATCH RECORD REVIEW ON 05/27/2019 TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED AND NO ISSUES WERE FOUND; ALL THE COMPONENTS FOR ASSEMBLY WERE CORRECT PER BILL OF MATERIALS (BOM) AND ALL THE TOOLING INFORMATION DOCUMENTED WAS ALSO CORRECT, UNDER INTERNATIONAL COMMODITY CODE (ICC) 175777 SYSTEM APPLICATION PRODUCT (SAP) MATERIAL ID 1000616 AND MANUFACTURING ORDER (B)(4). THE BATCH RECORD REVIEW SUPPORTS THAT THERE WERE NO DISCREPANCIES RELATED TO THE ISSUE REPORTED. RETURNED SAMPLE EVALUATION: NO PHOTOGRAPH ASSOCIATED WITH THIS CASE AND NO UNUSED RETURN SAMPLE WAS EXPECTED. CONCLUSION SUMMARY OF THE RELATED EVENT: BASED ON THE INVESTIGATION FINDINGS, THE ROOT CAUSE IDENTIFIED FOR THE ISSUE ¿WAFER OFF CENTER IN THE POUCH¿, REPORTED UNDER FAILURE MODE ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR¿ IS ATTRIBUTED TO: 1) MACHINE: THE INVESTIGATIVE PROCESS CONCLUDED THAT ALL THE MACHINERY/ TOOLING ITEMS USED IN THE MANUFACTURING PROCESS COMPLIED WHEN COMPARED AGAINST DRAWING AND PROCESS INSTRUCTION (PI) PI21-076 REV 37.0 SPECIFICATIONS; HOWEVER, AS PART OF THIS INVESTIGATION ON 10/MAY/2019, MAINTENANCE AND FACILITIES MANAGER AND SENIOR PROCESS ENGINEER PERFORMED A DEEP ASSESSMENT TO THE SUB-ASSEMBLY STATION OF THE CONVEX 1PC MACHINE IDENTIFYING THE CONDITIONS CAUSING THE WAFER OFF CENTER AND THE POSSIBLE SOLUTIONS TO CORRECT THE FAILURE. AS CONCLUSION OF THE ASSESSMENT, THE CHAIN THAT COORDINATES FLANGE LOADING STATION WITH ADHESIVE LOADING STATION NEEDED LUBRICATION AND THE CHAIN AND OTHER COMPONENTS OF THE MECHANISM NEEDED A CORRECTIVE MAINTENANCE. IN ORDER TO MAINTAIN THIS CONDITION IN THE LONG TERM, A DEEPER INTERVENTION OF THE MECHANISM IS REQUIRED. FURTHERMORE, REGARDING THE ROTARY TABLE STATION OF THE CONVEX 1PC MACHINE, IT WAS IDENTIFIED THAT DUE TO THE DEMANDS OF THE PROCESS, THE TOOLING (CUP POUCH URO 177A) USED TO MANUFACTURE THE URODRESS PRODUCT FAMILY REQUIRE A REDESIGN MODIFICATION TO REDUCE THE VARIABILITY WITHIN THE MANUFACTURING PROCESS. AS A CONTRIBUTOR FACTORS THE FOLLOWING OPPORTUNITIES FOR IMPROVEMENTS WERE IDENTIFIED: 2) PROCESS / METHOD INVESTIGATION: A) REFERENCE IN THE PI21-076; SECTION 8.0 ¿QUALITY INSTRUCTIONS¿ THE APPLICABLE QC TOOLING USED FOR PRODUCT¿S QUALITY INSPECTION PURPOSE OF POUCH URO 177A PRODUCTS FAMILY IN ORDER TO ADDRESS MANUFACTURING PERSONNEL THE USE OF THEM. B) IDENTIFICATION OF THE DEFECT DURING THE MANUFACTURING PROCESS, IT SHOULD BE CONSIDERED THE IMPLEMENTATION OF A JOB AID FOR PRODUCT CENTRALIZATION OUT OF SPECIFICATION. 3) PROCESS / METHOD / MEASUREMENT INVESTIGATION A) INTRODUCE QC TOOLING USED FOR PRODUCT¿S QUALITY INSPECTION PURPOSE IN THE CALIBRATION PROGRAM TO GUARANTEE MEASUREMENT ASSURANCE. 4) MANPOWER: A) THERE ARE NOT MANPOWER CAUSES ASSOCIATED WITH THE ROOT CAUSE; EVEN THOUGH, OPPORTUNITY TO IMPROVE THE LOADING OF THE ADHESIVE DISC DURING THE MANUFACTURING PROCESS WAS IDENTIFIED BY IMPLEMENTING A STANDARD WORK INSTRUCTION FOR CONVEX 1PC TO INDICATE MANUFACTURING OPERATORS HOW TO PROPERLY LOAD THE ADHESIVE DISC. NO ISSUES WERE IDENTIFIED FOR MATERIAL AND ENVIRONMENT INVESTIGATIONS. THE INVESTIGATION ASSOCIATED WITH RELATED EVENT HAS BEEN APPROVED AND IS COMPLETE. NO ADDITIONAL ACTION IS REQUIRED, AND THIS COMPLAINT WILL BE CLOSED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER. REPORTING SITE: 1049092. MANUFACTURING SITE: 9618003.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Additional Manufacturer Narrative · 1

BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE END USER REPORTED THE STARTER HOLE WAS OFF-CENTER. SHE FURTHER REPORTED SHE HAD DECREASED WEAR TIME, AND AS A RESULT HAS A TINY AREA OF DENUDED SKIN DUE TO THE FREQUENT APPLIANCE CHANGE. SHE USUALLY CHANGES HER APPLIANCE EVERY 7-10 DAYS BUT SHE HAD TO CHANGE THIS WITHIN 1 DAY. NO FURTHER INFORMATION OR PHOTOGRAPH WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
225117 ACTIVELIFE POUCH, COLOSTOMY EZQ CONVATEC DOMINICAN REPUBLIC INC 175777 8B00896 00768455181230

Patients

Seq Age Sex Outcome Treatment
1 Female