FDA Adverse Event Injury Summary report: N

PRECISION MONTAGE MRI

MDR report key: 8430718 · Received March 18, 2019

Report

Report Number
3006630150-2019-01148
Event Type
Injury
Date Received
March 18, 2019
Date of Event
January 10, 2019
Report Date
April 15, 2019
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
UDI-DI
08714729905943
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE EXPLANTED DEVICES WERE NOT RETURNED TO BSN AS THEY WERE KEPT BY THE MEDICAL FACILITY. A REVIEW OF THE MANUFACTURING DOCUMENTATION FOR THE IPG AND LEADS REVEALED THAT NO ANOMALIES OR DEVIATIONS POTENTIALLY RELATED TO THE EVENT OCCURRED DURING MANUFACTURING.

Description of Event or Problem · 0

A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF AND DISCOMFORT AT THE IPG AND LEAD SITE. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE.

Additional Manufacturer Narrative · 1

MODEL NUMBER/CATALOG NUMBER: SC-2218-50, SERIAL NUMBER: (B)(4), BATCH/LOT NUMBER: 5081488/5081495, MODEL/CATALOG DESCRIPTION: LINEAR ST LEAD KIT 50 CM.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF AND DISCOMFORT AT THE IPG AND LEAD SITE. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220963 PRECISION MONTAGE MRI STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1200 351570 08714729905943

Patients

Seq Age Sex Outcome Treatment
1 86 YR Required Intervention