FDA Adverse Event Malfunction Summary report: N

ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE

MDR report key: 8424098 · Received March 15, 2019

Report

Report Number
3001845648-2019-00110
Event Type
Malfunction
Date Received
March 15, 2019
Date of Event
February 14, 2019
Report Date
March 15, 2019
Manufacturer
COOK IRELAND LTD
Product Code
FCG
UDI-DI
00827002535855
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K) # K142688. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G.1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4) COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Additional Manufacturer Narrative · 0

PMA/510(K) # K142688. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING (B)(4). IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: (B)(4). THE ECHO-HD-19-C DEVICE OF LOT NUMBER C1544888 INVOLVED IN THIS COMPLAINT WAS RETURNED FOR EVALUATION, WITH THE ORIGINAL PACKAGING. THE PACKAGING WAS OPEN ON RECEIPT. WITH THE INFORMATION PROVIDED, A PHYSICAL EXAMINATION AND DOCUMENT BASED INVESTIGATION WAS CONDUCTED. THE DEVICE RELATED TO THIS OCCURRENCE UNDERWENT A LABORATORY EVALUATION ON THE (B)(6)2019. A DISTAL KINK WAS NOTED ON THE NEEDLE WHICH WAS CAUSING THE NOTCH TO CATCH ON THE SHEATH PREVENTING IT FROM RETRACTING FULLY. PRIOR TO DISTRIBUTION, ALL ECHO-HD-19-C DEVICES ARE SUBJECTED TO FUNCTIONAL CHECKS AND VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ECHO-HD-19-C OF LOT NUMBER C1544888 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS NOT PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1544888. THE NOTES SECTION OF THE INSTRUCTIONS FOR USE, WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO INSPECT THE DEVICE PRIOR TO USE : "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE". THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED AS THE CIRCUMSTANCES OF USE CANNOT BE REPLICATED IN THE LABORATORY. A POSSIBLE CAUSE COULD BE ATTRIBUTED TO A HARD LESION. COMPLAINT IS CONFIRMED AS THE FAILURE WAS VERIFIED IN THE LABORATORY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Description of Event or Problem · 0

THEY WERE NOT ABLE TO PULL THE NEEDLE FULLY BACK IN THE SHEAT SO THEY HAVE TO REMOVE THE NEEDLE WITH THE SHARP TIP OUTSIDE THE SHEATH.

Description of Event or Problem · 0

THIS FOLLOW UP REPORT IS BEING SUBMITTED DUE TO THE EVALUATION OF THE DEVICE INVOLVED IN THIS EVENT AND THE CONCLUSION OF THIS INVESTIGATION. THEY WERE NOT ABLE TO PULL THE NEEDLE FULLY BACK IN THE SHEAT SO THEY HAVE TO REMOVE THE NEEDLE WITH THE SHARP TIP OUTSIDE THE SHEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
217392 ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE FCG KIT, NEEDLE, BIOPSY FCG COOK IRELAND LTD G53585 C1544888 00827002535855

Patients

Seq Age Sex Outcome Treatment
1