ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Report
- Report Number
- 3001845648-2019-00108
- Event Type
- Malfunction
- Date Received
- March 15, 2019
- Date of Event
- July 24, 2018
- Report Date
- March 15, 2019
- Manufacturer
- COOK IRELAND LTD
- Product Code
- FCG
- UDI-DI
- 00827002535855
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PMA/510(K) # K142688. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G.1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4) COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.
PMA/510(K) # K142688. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING (B)(4). IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: (B)(4). COMPLAINT DEVICE WAS NOT RETURNED THEREFORE A DOCUMENT BASED REVIEW WILL BE PERFORMED. PRIOR TO DISTRIBUTION, ALL ECHO-HD-19-C DEVICES ARE SUBJECTED TO FUNCTIONAL CHECKS AND VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ECHO-HD-19-C OF LOT NUMBER C1295065 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS NOT PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1295065. THE NOTES SECTION OF THE INSTRUCTIONS FOR USE, WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO INSPECT THE DEVICE PRIOR TO USE : "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE" . THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE. THE FAILURE OF ADVANCEMENT/RETRACTION OF NEEDLE WAS CONCLUDED FROM THE AVAILABLE INFORMATION. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. A POSSIBLE ROOT CAUSE COULD BE ATTRIBUTED TO TORTUOUS ANATOMY REQUIRING FLEXED ENDOSCOPE POSITION AS INDICATED IN THE ADDITIONAL INFORMATION. THIS POTENTIALLY COULD HAVE CAUSED THE NEEDLE TO KINK RESULTING IN IT BEING UNABLE TO RETRACT INTO THE SHEATH. COMPLAINT IS CONFIRMED BASED ON THE CUSTOMER'S TESTIMONY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.
THE ECHOENDOSCOPY NEEDLE DID NOT INTERIORIZE INTO THE PROTECTIVE CATHETER WHEN IT WAS MANUALLY RETRACTED. IT WAS NECESSARY TO REMOVE THE DEVICE WITH THE EXPOSED NEEDLE. THERE WAS A RISK OF PERFORATING THE PATIENT AND THE DEVICE.
THIS FOLLOW UP REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION INTO THIS EVENT AND UPDATE TO THE CONCLUSION OF THIS INVESTIGATION. THE ECHO-ENDOSCOPY NEEDLE DID NOT INTERIORIZE INTO THE PROTECTIVE CATHETER WHEN IT WAS MANUALLY RETRACTED. IT WAS NECESSARY TO REMOVE THE DEVICE WITH THE EXPOSED NEEDLE. THERE WAS A RISK OF PERFORATING THE PATIENT AND THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 219171 | ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE | FCG KIT, NEEDLE, BIOPSY | FCG | COOK IRELAND LTD | G53585 | C1295065 | 00827002535855 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |