FDA Adverse Event Malfunction Summary report: N

ENFLOW

MDR report key: 8423041 · Received March 14, 2019

Report

Report Number
2050001-2019-00024
Event Type
Malfunction
Date Received
March 14, 2019
Date of Event
March 11, 2019
Report Date
March 12, 2019
Manufacturer
VYAIRE MEDICAL
Product Code
LGZ
UDI-DI
10885403286162
PMA / PMN Number
K130867
Removal / Correction Number
3008959444-3/13/19-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

VYAIRE FILE IDENTIFICATION: (B)(4). A RECENTLY CONDUCTED INTERNAL TESTING WITH THE RESULTS RECEIVED BY VYAIRE MEDICAL ON 11 MARCH 2019 INDICATES THAT THERE IS THE POTENTIAL TO ELUDE ALUMINUM FROM THE ENFLOW DISPOSABLE CARTRIDGE DURING INTRAVENOUS WARMING THERAPY WITH FLUID AND BLOOD SOLUTIONS. AS A RESULT, VYAIRE IS INITIATING AN IMMEDIATE GLOBAL RECALL OF THE ENFLOW DISPOSABLE CARTRIDGES PART NUMBER 980200EU & 980202EU. ANY ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER OR THROUGH VYAIRE'S INTERNAL INVESTIGATION WILL BE INCLUDED IN A FOLLOW-UP REPORT.

Description of Event or Problem · 1

A NEWS ARTICLE BY THE BMJ TITLED "ENFLOW FLUID WARMING DEVICE: WARNING OVER RISK OF ALUMINIUM TOXICITY" WAS PUBLISHED ON MARCH 11, 2019. THE ARTICLE MENTIONS A RECENT PAPER IN ANAESTHESIA THAT UNCOATED ALUMINIUM PLATES IN FLUID WARMING SYSTEMS, SUCH AS THOSE USED IN THE ENFLOW SYSTEM, YIELDED POTENTIALLY HARMFUL CONCENTRATIONS OF ALUMINIUM WHEN USING BALANCED ELECTROLYTE SOLUTIONS. LEAD RESEARCHER THORSTEN PERL, SENIOR CONSULTANT AT THE UNIVERSITY MEDICAL CENTRE (B)(6), TOLD THE BMJ THAT PATIENTS WITH IMPAIRED ALUMINIUM ELIMINATION FROM RENAL INSUFFICIENCY WERE AT RISK OF ADVERSE EVENTS. "AN EFFECTIVE COATING FOR ALUMINIUM CONTAINING MEDICAL DEVICES IN DIRECT CONTACT WITH FLUIDS SHOULD BE RECOMMENDED," HE SAID.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214820 ENFLOW ENFLOW 8482 DISPOSABLE IV FLUID/BLOOD WARMER CARTRIDGE WITH IV EXTENSION MODEL LGZ VYAIRE MEDICAL 980202EU 10885403286162

Patients

Seq Age Sex Outcome Treatment
1