CLEARVIEW SIMPLIFY D-DIMER
Report
- Report Number
- 8020026-2007-00002
- Event Type
- Death
- Date Received
- April 24, 2007
- Date of Event
- February 14, 2007
- Report Date
- April 23, 2007
- Manufacturer
- AGEN BIOMEDICAL LTD.
- Product Code
- GHH
- PMA / PMN Number
- k993276
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
IT IS NOT POSSIBLE TO DETERMINE WHETHER THE CLEARVIEW SIMPLIFY D-DIMER DEVICE USED TO TEST THE PATIENT IN QUESTION PRODUCED AN ACCURATE TEST RESULT OR NOT. IN ACCORDANCE WITH HOSPITAL PROCEDURE, THE PATIENT WAS SENT HOME AND THEREFORE, NO SCAN WAS PERFORMED WHICH WOULD HAVE CONFIRMED THE PRESENCE OF ANY CLOT AT THAT POINT IN TIME. NO BLOOD SAMPLE WAS AVAILABLE FOR TESTING BY ANOTHER METHOD. HOWEVER, THE WELLS SCORE WAS LOW WHICH INDICATES LOW PROBABILITY AND SUPPORTS THE RESULT OBTAINED WITH THE CLEARVIEW SIMPLIFY D-DIMER SUPPORT. IN ADDITION, THE OBSERVATION MADE BY THE PATIENT'S GP WAS THAT HE "APPEARED IN FINE HEALTH WITH NO SHORTNESS OF BREATH BUT HAD A COUGH." THIS INDICATES THE PATIENT WAS NOT SHOWING SIGNS AND SYMPTOMS OF DVT AT THE TIME OF PRESENTATION TO THE DOCTOR. A REVIEW OF OUR PRODUCT LABELING IS TO BE CARRIED OUT AND CHANGES WILL BE MADE IF DEEMED APPROPRIATE.
A MALE PATIENT WAS SEEN AT THE HOSPITAL AND ASSESSED FOR SUSPECTED DVT. THE WELLS PRE-TEST PROBABILITY SCORE WAS DESCRIBED BY THE HOSPITAL AS LOW (=1) INDICATING A LOW PROBABILITY OF DVT AND THE RESULT OBTAINED WITH THE CLEARVIEW SIMPLIFY D-DIMER TEST WAS NEGATIVE. THE SAMPLE USED IN THE CLEARVIEW SIMPLIFY D-DIMER TEST WAS VENOUS WHOLE BLOOD AND THE NURSE READ THE TEST AT THE INSTRUCTED READ TIME OF 10 MINUTES. UPON EXAMINATION, THE PATIENT'S LEG WAS NOT SWOLLEN. DVT WAS RULED OUT ON THE BASIS OF THE WELLS PRE-TEST PROBABILITY SCORE, CLEARVIEW SIMPLIFY D-DIMER RESULT AND EXAMINATION AS A RESULT, THE PATIENT WAS SENT HOME. THIS WAS IN ACCORDANCE WITH THE HOSPITAL'S PROTOCOL, THAT IS, A PATIENT PRESENTING AT THE HOSPITAL WITH A SUSPECTED DVT WOULD HAVE A WELLS PRE-TEST PROBABILITY SCORE. IF THIS IS LOW, THEN THE PATIENT IS TESTED WITH THE CLEARVIEW SIMPLIFY D-DIMER TEST AND IF THE RESULT IS NEGATIVE, THE PATIENT IS SENT HOME. IF THE CLEARVIEW SIMPLIFY D-DIMER TEST IS POSITIVE, THEN THE PATIENT IS SENT FOR A SCAN. THE PATIENT VISITED HIS GP AND HIS CONDITION WAS DESCRIBED BY HIS GP AS "APPEARED IN FINE HEALTH, NO SHORTNESS OF BREATH BUT HAD A COUGH." THE GP PRESCRIBED ANTIBIOTICS TO TREAT THE SUSPECTED UPPER RESPIRATORY TRACT INFECTION. THE PATIENT DIED THREE DAYS LATER. THE PATIENT WAS SAID (BY THE CORONER) TO HAVE HAD A MASSIVE EMBOLI IN BOTH MAIN PULMONARY ARTERIES ORIGINATING FROM A DVT. THE PATIENT DIED 8 DAYS AFTER TESTING WITH CLEARVIEW SIMPLIFY D-DIMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLEARVIEW SIMPLIFY D-DIMER | GHH | GHH | AGEN BIOMEDICAL LTD. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Death |