LIGHTSHEER (ALL MODELS)
Report
- Report Number
- 2914019-2007-00013
- Event Type
- Other
- Date Received
- April 25, 2007
- Date of Event
- March 23, 2007
- Report Date
- April 25, 2007
- Manufacturer
- LUMENIS, LTD.
- Product Code
- GEX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
Narratives
OTHER ROOT CAUSE OF THE INCIDENT APPEARS TO HAVE BEEN DEVICE OUTPUT FROM THE SR HEAD THAT WAS ELEVATED ( OUT OF SPECIFICATION). PER THE LUMENIS CUSTOMER ENGINEER CE, THE IPL CALIBRATION ADAPTOR FOR THE USER'S POWER METER WAS DAMAGED. THE GLASS WINDOW WAS SCARRED, WHICH WOULD CAUSE THE IPL TREATMENT HEAD TO CALIBRATE TOO HIGH DUE TO POOR LIGHT TRANSMISSION TO THE POWER METER. DAMAGE APPEARS TO BE 2 CIRCULAR BURNS 135 DEGREES APART, WHICH SCARRED ABOUT 1/3 OF THE TOTAL SURFACE AREA OF THE WINDOW. WHEN THE DEVICE OPERATOR CALIBRATED THE SR TREATMENT HEAD THROUGH THE DAMAGED POWER METER GLASS, A HIGH CALIBRATION OCCURRED. THE CE REPLACED THE ENTIRE POWER METER ASSEMBLY, SYSTEM RETURNED TO NORMAL AFTER RECALIBRATING FOR IPL OPERATION. SYSTEM PASSED OPERATIONAL AND SAFETY CHECKS. DURING INVESTIGATION WITH THE CUSTOMER, NO CAUSE (SUCH AS FOREIGN MATERIAL ON THE ENERGY METER GLASS) COULD BE ESTABLISHED FOR THE DAMAGE TO THE ENERGY METER GLASS. PER THE CUSTOMER, THE DEVICE WAS FREE OF DEBRIS AT THE TIME OF CALIBRATION. CUSTOMER WAS RECOMMENDED TO PERFORM A VISUAL INSPECTION OF THE ENERGY METER GLASS PRIOR TO EACH CALIBRATION AND - IF ANY CHANGE IN APPEARANCE OR DAMAGE TO THE GLASS ARE NOTED - TO SUSPEND USE OF THE DEVICE AND CALL LUMENIS SERVICE FOR EVALUATION.
PER THE CUSTOMER, A PATIENT WAS BURNED IN ASSOCIATION WITH IPL HAIR REMOVAL TREATMENT TO THE RIGHT SIDE OF THE FACE. PER THE CUSTOMER, THE INCIDENT OCCURRED IN 2007. THE STAFF AESTHETICIAN LATER TESTED THE DEVICE ON HERSELF AND ALSO WAS BURNED. THE INCIDENT WAS REPORTED TO THE LUMENIS CE DURING A PERIOIDIC MAINTENANCE (PM) SERVICE VISIT. SERVICE ANALYSIS OF THE SAFETY COMPLAINT WAS PERFORMED DURING THE PM CALL WHILE THE CE WAS ONSITE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGHTSHEER (ALL MODELS) | OTHER | GEX | LUMENIS, LTD. | QUANTUM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |