FDA Adverse Event Malfunction Summary report: N

UNIVERSAL SLINGB 450 QRH

MDR report key: 8420299 · Received March 14, 2019

Report

Report Number
8030916-2019-00018
Event Type
Malfunction
Date Received
March 14, 2019
Date of Event
February 15, 2019
Report Date
February 15, 2019
Manufacturer
LIKO AB
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT PROVIDED A PICTURE OF THE BROKEN COMPOSITE SLING BAR HOOK. THE MOST PROBABLE CAUSE FOR THE COMPOSITE HOOK TO BREAK IS THE LOAD BEING APPLIED ASYMMETRICALLY ON ONLY ONE SIDE OF THE SLING BAR. TESTS PERFORMED ON UNIVERSAL SLING BARS HAVE SHOWN THAT WHEN THE LOAD IS APPLIED TO BOTH SLING BAR HOOKS, AS PER THE INTENDED USE, THE COMPOSITE HOOK CAN WITHSTAND > 780 KG BEFORE BREAKAGE. IN THE INSTRUCTION GUIDE FOR UNIVERSAL SLING BARS (7EN160185 REV. 4), THE SAFETY INSTRUCTION SECTION STATES: BEFORE LIFTING, ALWAYS MAKE SURE THAT THE SLING¿S STRAP LOOPS ARE CORRECTLY FASTENED TO THE SLING BAR HOOKS WHEN THE SLING STRAP IS EXTENDED, BUT BEFORE THE PATIENT IS LIFTED FROM THE UNDERLYING SURFACE. FOR SAFETY, LOAD MUST BE APPLIED TO ALL SLING BAR HOOKS ON THE SLING BAR DURING THE LIFT! A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS DEVICE. THE FACILITY HAD A THIRD PARTY COMPANY PERFORM PREVENTATIVE MAINTENANCE ON THIS DEVICE. THE ACCOUNT WAS SENT A NEW SLING BAR TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING UNIVERSAL SLINGBAR 450 BROKE. THE LIFT WAS LOCATED AT THE ACCOUNT AT THE TIME OF THE INCIDENT. THERE WAS NO ALLEGATION OF PATIENT OR USER INJURY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
216390 UNIVERSAL SLINGB 450 QRH NON-AC POWERED PATIENT LIFT FSA LIKO AB 3156085

Patients

Seq Age Sex Outcome Treatment
1