FREESTYLE FLASH
MDR report key: 841798
·
Received October 13, 2006
Report
- Report Number
- 2954323-2006-01043
- Event Type
- Malfunction
- Date Received
- October 13, 2006
- Date of Event
- August 4, 2006
- Report Date
- October 13, 2006
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- LFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE METER WAS CONFIRMED TO HAVE EXHIBITED MEMORY OVERWRITE. TESTED RETURNED UNIT. NO MFG PARAMETERS FOUND OUT OF SPEC AND CIG PANASONIC BATTERY VOLTAGE WAS MEASURED AT 2.975V. DID NOT OBSERVE BATTERY ICON OR BOOKLET ICON WHILE STRIP WAS INSERTED. HOWEVER, UOM WAS SELECTABLE AND WAS REC'D AT MMOL/L. 0300, 0800, 0204 AND 0015 ERRORS WERE OBSERVED IN THE ERROR LOG INDICATING BATTERY DROOP. METER IS OPERATIONAL.
Description of Event or Problem · 1
CUSTOMER REPORTS HAVING INCORRECT UNIT OF MEASURE ON HER ADC METER. THE CUSTOMER ALSO REPORTS SEEING A BATTERY ICON AND HAVING OTHER SETTINGS OF THE ADC METER CHANGED WHICH INDICATES A POTENTIAL MEMORY OVERWRITE ISSUE. NO DEATH OR SERIOUS INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FLASH | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC. | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |