PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 3004209178-2019-66228
- Event Type
- Malfunction
- Date Received
- March 12, 2019
- Date of Event
- October 30, 2018
- Report Date
- March 12, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 00763000090203
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
NO FAILED BATT TEST ALARMS NOTED DURING TESTING. UNIT PASSED DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELFTEST. TOGGLED ON/OFF AND INCREASED/DECREASED VOLUME WITH THE AUDIO FEATURE FUNCTIONING PROPERLY. NO AUDIO/VIBRATE/ABSENCE OF ALARM ANOMALIES NOTED DURING TESTING. PUMP ERROR 54 WAS PRESENT IN THE PUMP TRACE DOWNLOAD DUE TO SOFTWARE ERROR (ESF 3010978). PUMP ERROR WAS PRESENT IN THE PUMP TRACE DOWNLOAD, PROBLEM ISOLATED TO ELECTRONIC BOARD STACK.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED MULTIPLE PUMP ERRORS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER RECEIVED REQUEST TO REWIND.THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACKUP PLAN. THE DEVICE WAS EXPECTED TO BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 205302 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG2K1KX | 00763000090203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |