AQUABEAM SYSTEM
Report
- Report Number
- 3012977056-2019-00008
- Event Type
- Malfunction
- Date Received
- March 11, 2019
- Date of Event
- February 12, 2019
- Report Date
- May 21, 2021
- Manufacturer
- PROCEPT BIOROBOTICS
- Product Code
- PZP
- PMA / PMN Number
- DEN170024
- Removal / Correction Number
- 3012977056-030619-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
H. 11 CORRECTED DATA: FOR CORRECTED DATA, PLEASE REFER TO SECTION H.1 TYPE OF REPORTABLE EVENT, FOLLOW-UP #1 WAS INCORRECTLY SUBMITTED AS "SERIOUS INJURY".
H.10. [X] ADDITIONAL MANUFACTURER NARRATIVE. THE AQUABEAM HANDPIECE WAS RETURNED FOR INVESTIGATION. THE AQUABEAM HANDPIECE WAS CONNECTED TO AN AQUABEAM SYSTEM AND A TEST AQUABLATION TREATMENT RUN WAS PERFORMED; A PERSISTENT E22 ERROR WAS OBSERVED. THE TEST TREATMENT PROCEDURE WAS ABLE TO BE COMPLETED. THE HANDPIECE WAS DISCONNECTED, THE OUTER SHELL WAS REMOVED, AND THE HANDPIECE WAS CONNECTED AGAIN TO THE AQUABEAM SYSTEM AND A TEST TREATMENT RUN WAS PERFORMED WITHOUT THE OUTER SHELL. A LEAK WAS OBSERVED AT THE P1080 PROBE TUBE TO P1160 PUMP CARTRIDGE DISTAL CRIMP LASER WELD. THE WELD LEAK RESULTED IN SALINE LEAKING ON THE SENSOR BOARD, WHICH CAUSED AN R POSITION LOST ERROR. THE ERROR IS DISPLAYED TO THE USER AS AN E22 ERROR ON THE AQUABEAM CPU. THE INVESTIGATION CONFIRMED THE WELD LEAK LIKELY CAUSED THE REPORTED E22 ERROR. THE REPORTED EVENT WAS CONFIRMED DURING THE INVESTIGATION. ROOT CAUSE INVESTIGATION INTO THIS ISSUE DETERMINED THAT THE MICRO-CRACK THAT CAN DEVELOP IN A WELD JOINT WITHIN THE HANDPIECE, AND CAUSE THE FAILURE MODE, IS DUE TO VARIABILITY IN A SINGLE PASS WELDING PROCESS. THE CONTRIBUTING FACTORS TO THE VARIABILITY WERE IDENTIFIED THROUGH A FISH BONE ANALYSIS AND ARE LISTED BELOW. 1. WELD PLACEMENT. 2. COMPONENT GAP. 3. WELD DESIGN. 4. WELD INSPECTION. 5. MATERIAL COMPATIBILITY. 6. INCREASE IN PRODUCTION VOLUME. MULTIPLE LONG-TERM CORRECTIVE AND PREVENTIVE ACTIONS WERE ESTABLISHED TO FURTHER ADDRESS THE UNDERLYING FACTORS IDENTIFIED. THIS EVENT IS PART OF RECALL 3012977056-030619-001-R. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
DURING AN AQUABLATION PROCEDURE, E22 (R MOTION ERROR) MESSAGE OCCURRED TWICE. AFTER 60 MINUTES OF TROUBLE SHOOTING WITH TWO MALFUNCTIONED AQUABEAM HANDPIECES REPLACED, THE PROCEDURE WAS COMPLETED WITH NO PATIENT INJURY REPORTED. THIS REPORT IS FOR ONE OF THE MALFUNCTIONED HANDPIECES WITH LOT# 18C00681.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 202687 | AQUABEAM SYSTEM | FLUID JET REMOVAL SYSTEM | PZP | PROCEPT BIOROBOTICS | 310301 | 18C00681 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |