V60 VENTILATOR
Report
- Report Number
- 2031642-2019-01406
- Event Type
- Malfunction
- Date Received
- March 11, 2019
- Report Date
- February 14, 2019
- Manufacturer
- RESPIRONICS CALIFORNIA, INC
- Product Code
- MNT
- UDI-DI
- 00884838020054
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
DATE OF REPORT: 30MAY2019. DATE REC¿D BY MFR: 29MAY2019. THE REPORTED ISSUE PROXIMAL PRESSURE SENSOR AUTO ZERO FAILED THE PHILIPS FIELD SERVICE ENGINEER (FSE) RETURNED THE DATA ACQUISITION PRINTED CIRCUIT BOARD ASSEMBLY (DAQ PCBA). THE FSE ALSO REPLACED SOLENOIDS 3 AND 4 BUT DID NOT RETURN THEM FOR INVESTIGATION. A DAQ PCBA WAS RETURN FOR ANALYSIS. A VISUAL INSPECTION OF THE DATA ACQUISITION PRINTED CIRCUIT BOARD ASSEMBLY (DAQ PCBA) REVEALED NO ANOMALIES. THE RETURNED DAQ PCBA WAS INSTALLED INTO THE FAILURE INVESTIGATION (FI) TEST VENTILATOR IN AN ATTEMPT TO DUPLICATE THE REPORTED PROBLEM PROXIMAL PRESSURE SENSOR AUTO ZERO FAILED. THE FI TECHNICIAN WAS UNABLE TO DUPLICATE THE REPORTED FAILURE IN THE RETURNED DAQ PCBA. SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
DATE OF EVENT: (B)(6) 2019. DATE OF REPORT: 11MARCH2019. PHILIPS FIELD SERVICE ENGINEER (FSE) CONFIRMED PROXIMAL PRESSURE SENSOR AUTOZERO FAILED (ERROR CODE 110A) SEEN AT SEVERAL INSTANCES IN THE DEVICE HISTORY REPORT. THE FSE REPLACED TWO OF THE SOLENOID VALVES AND THE DIGITAL TO ANALOG CONVERTER BOARD. THE UNIT WAS READY FOR PATIENT USE.
THE CUSTOMER REPORTED THAT THE SYSTEM WAS MALFUNCTIONING. THERE WAS NO PATIENT OR USER HARM REPORTED. THE EVENT DATE WAS NOT SPECIFIED; ESTIMATE USED.