FDA Adverse Event Other Summary report: N

PLATO BPS

MDR report key: 84096 · Received April 14, 1997

Report

Report Number
1121753-1997-00008
Event Type
Other
Date Received
April 14, 1997
Date of Event
January 24, 1996
Report Date
April 11, 1997
Manufacturer
NUCLETRON CORP.
Product Code
IWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

AN INCORRECT RECONSTRUCTION IS POSSIBLE WHEN THE DIGITIZER IS USED AS THE IMPLANT RECONSTRUCTION METHOD. THE USER IS REQUESTED TO DIGITIZE TWO DIAGONAL CORNERS OF A RECTANGLE, THUS DEFINING AN AREA FROM WHICH THE RELEVANT ANATOMICAL AND IMPLANT INFO IS OBTAINED. DURING THE SETUP PROCEDURE, "ORIGIN" AND "POINT ON POS. X-AXIS" REQUIRE ENTRY TO POSITION THE FILM ON THE DIGITIZER. IF THESE PARAMETERS ARE NOT DEFINED, A WARNING MESSAGE IS GENERATED: "WARNING, CROSSWIRES NOT ORIENTED!" IF THE USER IGNORES THE WARNING AND PRESSES THE "OK" BUTTON, THE PROGRAM ALLOWS THE USER TO CONTINUE. THIS CAN RESULT IN AN INCORRECT RECONSTRUCTION OF THE IMPLANT IF THE CROSSWIRES ARE NOT ORIENTED TO THE CENTER OF THE AREA DEFINED BY THE DIAGONAL CORNERS OF A RECTANGLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLATO BPS TREATMENT PLANNING SYSTEM IWB NUCLETRON CORP. PLATO-BPS V. 13.1 *

Patients

Seq Age Sex Outcome Treatment
1 *