FDA Adverse Event Malfunction Summary report: N

EVIS EXERA LLL COLONOVIDEOSCOPE

MDR report key: 8406876 · Received March 11, 2019

Report

Report Number
8010047-2019-01237
Event Type
Malfunction
Date Received
March 11, 2019
Date of Event
January 31, 2019
Report Date
April 8, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDF
UDI-DI
04953170305177
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) BUT WAS RETURNED TO OLYMPUS FRANCE (OFR). OFR SENT THE DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. THE ADDITIONAL MICROBIOLOGICAL TESTING INDICATED NO MICROBIAL GROWTH FOR THE ALL CHANNELS OF THE DEVICE. THEREFORE, THE TESTING RESULT CLEARED THE FRENCH GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC FOR EVALUATION. OMSC REVIEWED THE MANUFACTURING HISTORY OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MULTIPLE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING AMOUNT OF UNSPECIFIED MICROBES WERE DETECTED WITH THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. [FIRST TIME; (B)(6) 2019]: 17 CFU / 100ML, [SECOND TIME; (B)(6) 2019]: 16 CFU / 100ML, [THIRD TIME; (B)(6) 2019]: 16 CFU / 100ML. THE SUBJECT DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE S3/S4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
200660 EVIS EXERA LLL COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF OLYMPUS MEDICAL SYSTEMS CORP. CF-H190I 04953170305177

Patients

Seq Age Sex Outcome Treatment
1