FDA Adverse Event Malfunction Summary report: N

POWERFLEX P3 PTA BALLOON CATHETER

MDR report key: 840483 · Received February 26, 2007

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
9610978-2007-00113
Event Type
Malfunction
Date Received
February 26, 2007
Date of Event
February 9, 2007
Report Date
February 9, 2007
Manufacturer
CORDIS EUROPA, N.V.
Product Code
LIT
PMA / PMN Number
K981407
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED TO CORDIS FOR ANALYSIS. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND SHOWED THAT THIS LOT OF PRODUCTS (R0306101) MET ALL REQUIREMENTS PER THE APPLICABLE MANUFACTURING QUALITY PLAN (MQP). THIS REVIEW WAS EXTENDED TO SUBASSEMBLY PART NUMBER E7145121 WITH LOT NUMBER 0102060015 AND 0102060957. THIS REVIEW CONFIRMED THAT SUBASSEMBLIES MET ALL REQUIREMENTS PER THE APPLICABLE MQP AS WELL, INCLUDING THE BURST TEST VALUES. IN THIS CASE, THERE IS NOT ENOUGH INFORMATION TO DRAW A DEFINITIVE CONCLUSION BETWEEN THE DEVICE AND THE REPORTED EVENT, HOWEVER, VESSEL CHARACTERISTICS MAY HAVE BEEN A CONTRIBUTING FACTOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A LEAKAGE IN THE BALLOON AND THE BALLOON WOULD NOT FULLY INFLATE. THE INTENDED PROCEDURE WAS ANGIOPLASTY OF THE ILIAC ARTERY. THE VESSEL HAD MILD CALCIFICATION, MILD TORTUOSITY AND A 60% STENOSIS. THE POWERFLEX P3 BALLOON CATHETER WAS ADVANCED TO THE LESION AND INFLATED WITH AN INDEFLATOR, HOWEVER, THE BALLOON WOULD ONLY INFLATE TO 4 ATMOSPHERES. THE CATHETER WAS REMOVED AND A LEAKAGE IN THE BALLOON WAS NOTED. NO ADDITIONAL PROCEDURAL DETAILS WERE PROVIDED. IT IS UNKNOWN IF DIFFICULTY WAS EXPERIENCED DURING ADVANCEMENT OF THE BALLOON CATHETER TO THE LESION OR WHILE CROSSING THE LESION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERFLEX P3 PTA BALLOON CATHETER PTA CATHETERS (LIT) LIT CORDIS EUROPA, N.V. NA R0306101

Patients

Seq Age Sex Outcome Treatment
1 *