FDA Adverse Event Malfunction Summary report: N

BD MICRO-FINE IV INSULIN SYRINGE

MDR report key: 8401471 · Received March 7, 2019

Report

Report Number
1920898-2019-00232
Event Type
Malfunction
Date Received
March 7, 2019
Date of Event
January 28, 2019
Report Date
March 14, 2019
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
30382903294207
PMA / PMN Number
K170386
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: CUSTOMER RETURNED PHOTOS OF A 1ML SYRINGE. CUSTOMER STATES THAT THE NEEDLES BREAK OR "CROOKS" WHEN INJECTING INTO THE SKIN OR WHEN RECAPPING THE NEEDLE. THE ATTACHED PHOTOS WERE EXAMINED AND NO BENT OR BROKEN CANNULA WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7121840. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE TWO (2) NOTIFICATIONS [(B)(4)] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THAT THE NEEDLES BREAK OR "CROOKS" WHEN INJECTING INTO THE SKIN OR WHEN RECAPPING THE NEEDLE.

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 7240669; MEDICAL DEVICE EXPIRATION DATE: 2022-10-31; DEVICE MANUFACTURE DATE: 2017-08-28. MEDICAL DEVICE LOT #: 7121840; MEDICAL DEVICE EXPIRATION DATE: 2022-05-31; DEVICE MANUFACTURE DATE: 2017-05-01. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THE BD MICRO-FINE¿ IV INSULIN SYRINGE THAT THE NEEDLES BREAK OR "CROOKS" WHEN INJECTING INTO THE SKIN OR WHEN RECAPPING THE NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194259 BD MICRO-FINE IV INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE SEE H.10 30382903294207

Patients

Seq Age Sex Outcome Treatment
1 Other