FDA Adverse Event Malfunction Summary report: N

BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE

MDR report key: 8400206 · Received March 7, 2019

Report

Report Number
9616656-2019-00216
Event Type
Malfunction
Date Received
March 7, 2019
Date of Event
February 8, 2019
Report Date
March 21, 2019
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
UDI-DI
00382903296057
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: CUSTOMER RETURNED (1) OPEN 5MM, 30G BD AUTOSHIELD DUO SAMPLES FROM LOT # 8081867. CUSTOMER STATES THAT DURING THE APPLICATION OF INSULIN, WITH THE CORRECT TECHNIQUE, AFTER 10 SEC. PRESENTED EXTRAVASATION OF INSULIN IN OSTIUM. THE RETURNED PEN NEEDLE WAS RETURNED NOT ACTIVATED. THE SAMPLE WAS TESTED AND WAS ABLE TO EXPEL PROPERLY AND ACTIVATE PROPERLY WITHOUT ANY OBSERVED DEFECTS. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHEN USING A BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE DURING APPLICATION OF INSULIN THERE WAS EXTRAVASATION OF INSULIN IN OSTIUM. THE REPORT IS AS FOLLOWS, "DURING THE APPLICATION OF INSULIN, WITH THE CORRECT TECHNIQUE, AFTER 10 SEC. PRESENTED EXTRAVASATION OF INSULIN IN OSTIUM."

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN USING A BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE DURING APPLICATION OF INSULIN THERE WAS EXTRAVASATION OF INSULIN IN OSTIUM. THE REPORT IS AS FOLLOWS, "DURING THE APPLICATION OF INSULIN, WITH THE CORRECT TECHNIQUE, AFTER 10 SEC. PRESENTED EXTRAVASATION OF INSULIN IN OSTIUM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195624 BD AUTOSHIELD¿ DUO SAFETY PEN NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 8081867 00382903296057

Patients

Seq Age Sex Outcome Treatment
1 Other