C-VUE
Report
- Report Number
- 1034196-2007-00001
- Event Type
- Injury
- Date Received
- April 12, 2007
- Date of Event
- March 24, 2007
- Report Date
- April 12, 2007
- Manufacturer
- UNILENS CORP., USA
- Product Code
- HPX
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
EVAL OF THE RETURNED PORTION OF THE LENS WAS INCONCLUSIVE. THE NATURE OF THE FRACTURE AS SEEN UNDER VISIBLE AND POLARIZED LIGHT WAS OF A PROPAGATED FRACTURE ACROSS THE OPTIC, WITH NO APPARENT STRESS FRACTURES INDICATING A POINT OF IMPACT OR DAMAGE. NO CONCLUSIONS CAN BE DRAWN CONCERNING THE DAMAGE THAT OCCURRED TO THE LENS TO CAUSE IT TO BREAK INTO PIECES.
ON 26 MAR 2007, NOTIFIED BY SALES REP THAT EYE CLINIC HAD A PT REPORT A CVUE GP CONTACT LENS BROKE IN THE EYE, AND WAS TREATED AT THE ER. WHEN CONTACTED LATER IN THE DAY, THE CONTACT LENS ASSOCIATE REVIEWED THE ER REPORT: ON SATURDAY MORNING, THE PT NOTED HIS RIGHT EYE BEGAN TO HURT, AND COULD NOT REMOVE HIS CONTACT. HE PRESENTED AT THE ER AND WAS SEE BY AN ATTENDING PHYSICIAN. ONLY ONE PORTION OF THE LENS REMAINED IN THE EYE, AND AFTER SEARCHING AND FLUSHING THE EYE, THE REMAINING PORTION WAS RETRIEVED. THE PT WAS DIAGNOSED WITH A CORNEAL ABRASION, WAS PROVIDED WITH VIGAMOX ANTIBIOTIC FOR APPLICATION 3X DAILY, INSTRUCTED TO LEAVE HIS LENSES OUT AND TO SEE HIS EYECARE PRACTITIONER ON MONDAY. THE PT'S VISUAL ACUITY WITH SPECTACLE CORRECTION WAS 20/60 IN THE RIGHT EYE AT THE CONCLUSION OF THE ER VISIT. THREE DAYS LATER,ADVISED BY A PHONE MAIL THAT THE PT HAD SEEN THE EYECARE PRACTITIONER AND THE EYE WAS FOUND IN EXCELLENT CONDITION WITH NO RESIDUAL EFFECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | C-VUE | RIGID GAS PERMEABLE CONTACT LENS | HPX | UNILENS CORP., USA | CVUE GP | 907107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |