FDA Adverse Event Malfunction Summary report: N

BD MICRO-FINE ULTRA¿ INSULIN PEN NEEDLE

MDR report key: 8397080 · Received March 6, 2019

Report

Report Number
9616656-2019-00215
Event Type
Malfunction
Date Received
March 6, 2019
Date of Event
February 15, 2019
Report Date
March 26, 2019
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: 53 SEALED AND INTACT 31G X 8MM PEN NEEDLE SAMPLES WERE RETURNED FROM LOT. NO. 8030906, CAT. NO. 325105. A CLOG TEST WAS CARRIED OUT ON FIFTY THREE SAMPLES AND NO ISSUES WERE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. NO ISSUES WERE OBSERVED WITH THE RETURNED SAMPLES THEREFORE NO ROOT CAUSE CAN BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SOME OF THE BD MICRO-FINE ULTRA¿ INSULIN PEN NEEDLE ARE CLOGGED.

Additional Manufacturer Narrative · 1

A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SOME OF THE BD MICRO-FINE ULTRA¿ INSULIN PEN NEEDLE ARE CLOGGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186853 BD MICRO-FINE ULTRA¿ INSULIN PEN NEEDLE PEN NEEDLE FMI BECTON DICKINSON AND CO. 8030906

Patients

Seq Age Sex Outcome Treatment
1 Other