FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 839584
·
Received April 12, 2007
Report
- Report Number
- 1419798-2007-00013
- Event Type
- Injury
- Date Received
- April 12, 2007
- Date of Event
- March 20, 2007
- Report Date
- April 11, 2007
- Manufacturer
- KAVO AMERICA
- Product Code
- EKK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
LIGHT RESIDUE, BACK UP CAP STUCK DUE TO DENTED HEAD PUTTING PRESSURE ON THE BEARINGS CAUSING THE HEAD TO HEAT UP. MAINTENANCE INFORMATION WAS REVIEWED WITH THE DOCTOR AND MAINTENANCE SHEETS WERE SENT.
Description of Event or Problem · 1
THE HANDPIECE BECAME HOT AND BURNED THE PATIENT LIP. THE BURNS WERE THE SIZE OF THE BACK CAP. OINTMENT WAS APPLIED TO THE BURN.