FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 839572
·
Received April 12, 2007
Report
- Report Number
- 1419798-2007-00014
- Event Type
- Injury
- Date Received
- April 12, 2007
- Date of Event
- March 21, 2007
- Report Date
- April 11, 2007
- Manufacturer
- KAVO AMERICA
- Product Code
- EKK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
HEAD IS DAMAGED DUE TO MULTIPLE DENTS BEARINGS/ GEARS ARE WORN AND GRITTY IN DRIVE ASSEMBLY AND SHAFT HAD HEAVY DEBRIS. MAINTENANCE INFORMATION WAS REVIEWED WITH THE DOCTOR AND MAINTENANCE SHEETS WERE SENT.
Description of Event or Problem · 1
THE HANDPIECE BECAME HOT AND BURNED THE PATIENT LIP. THE BURNS WERE SLIGHTLY LARGER THAN HANDPIECE HEAD. OINTMENT WAS APPLIED TO THE BURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | OPERATIVE DENTAL UNIT ACCESSORY | EKK | KAVO AMERICA | 25LPA | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention |