FDA Adverse Event Malfunction Summary report: N

BD 50ML SYRINGE

MDR report key: 8395665 · Received March 6, 2019

Report

Report Number
2243072-2019-00419
Event Type
Malfunction
Date Received
March 6, 2019
Date of Event
February 18, 2019
Report Date
April 23, 2019
Manufacturer
BECTON DICKINSON
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: 1 SAMPLE WAS RETURNED TO SBDM. FROM INVESTIGATIONS, THE LIKELY CAUSE IS DUE TO SYRINGE ASSEMBLY PROCESS, SBDM ASSUME THAT THE FM IN THE BARREL WAS OCCURRED WHILE THE PLUNGER WAS FEEDING TO SYRINGE ASSEMBLY MACHINE. THE PLUNGER WAS NOT MOVING TO INDEX OF SYRINGE ASSEMBLY MACHINE PROPERLY, THUS CAUSING THE PLUNGER TO BREAK. THE BROKEN PIECE MOST LIKELY WAS PUSHED BY AIR INTO THE BARREL, WHEN THE ASSEMBLY MACHINE WAS STARTING OR STOPPING. HOUSE SAMPLE INSPECTION: SBDM INSPECTED 30 PCS EACH FOR 3 LOTS (LOT 1810013, 1810224 AND 1812062, NO ABNORMALITY WAS OBSERVED. DEVICE HISTORY R REVIEW: SBDM REVIEWED THE MANUFACTURING RECORD FOR LOT 1810224, NO ABNORMALITY WAS OBSERVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD¿ 50ML SYRINGE HAD FOREIGN MATTER. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(4). THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AND THE (B)(4) FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD¿ 50ML SYRINGE HAD FOREIGN MATTER. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186961 BD 50ML SYRINGE SYRINGE FMF BECTON DICKINSON 1810224

Patients

Seq Age Sex Outcome Treatment
1 Other