FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 839563
·
Received April 12, 2007
Report
- Report Number
- 1419798-2007-00015
- Event Type
- Injury
- Date Received
- April 12, 2007
- Date of Event
- March 26, 2007
- Report Date
- April 11, 2007
- Manufacturer
- KAVO AMERICA
- Product Code
- EKK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
HEAD IS DAMAGED DUE TO DENT AT BACK END CAP. BEARING AND GEARS ARE WORN. GRITTY IN DRIVE ASSEMBLY. SHAFT AND AXLE HAVE HEAVY DEBRIS. MAINTENANCE INFORMATION WAS REVIEWED WITH THE OFFICE AND MAINTENANCE SHEETS WERE SENT.
Description of Event or Problem · 1
THE HANDPIECE BECAME HOT AND BURNED THE PATIENT LIP. OINTMENT WAS APPLIED TO THE BURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | OPERATIVE DENTAL UNIT ACCESSORY | EKK | KAVO AMERICA | 25LPA | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention |