Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE SUBJECT HAS OUT OF RANGE IMPEDANCES AT 3V FOR THE RIGHT HEMISPHERE: 0 AND 1 - 7123 0 AND 2 - 8294 0 AND 3 - 8360 C AND 0 - 5521 NO FURTHER COMPLICATIONS WERE REPORTED.

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 3389, SERIAL#: (B)(4), IMPLANTED:(B)(6) 2018, PRODUCT TYPE: LEAD. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 3389, SERIAL/LOT #: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE MAIN COMPONENT OF THE SYSTEM AND OTHER APPLICABLE COMPONENTS ARE: PRODUCT ID: 3389, LOT# VAILZXX, IMPLANTED: (B)(6) 2018, PRODUCT TYPE: LEAD. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT ID: 3389, LOT# VAILZXX, IMPLANTED: (B)(6) 2018, PRODUCT TYPE: LEAD. EVENT DATE CORRECTED. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT'S WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) IN A CLINICAL STUDY. IT WAS REPORTED THE SUBJECT HAD OUT OF RANGE IMPEDANCES AT 0.7V FOR THE RIGHT HEMISPHERE LEAD ON (B)(6) 2019. C AND 0/8 5171 C AND 2/10 2169 C AND 3/11 2412 0/8 AND 1/9 6398 0/8 AND 2/10 6933 0/8 AND 3/11 7416 1/9 AND 3/11 4519 2/10 AND 3/11 4035 NO KNOWN ACTIONS WERE TAKEN. THE ISSUE WAS ONGOING AT THE TIME OF THE EVENT.

Description of Event or Problem · 1

NO ADDITIONAL INFORMATION WAS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188868 ACTIVA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MEDTRONIC PUERTO RICO OPERATIONS CO. 37603

Patients

Seq Age Sex Outcome Treatment
1 54 YR