FDA Adverse Event Malfunction Summary report: N

BD CATHENA¿ 22GX1.00IN STRAIGHT BC

MDR report key: 8392426 · Received March 5, 2019

Report

Report Number
8041187-2019-00200
Event Type
Malfunction
Date Received
March 5, 2019
Date of Event
February 13, 2019
Report Date
March 27, 2019
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
UDI-DI
00382903868063
PMA / PMN Number
UNKNOWN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: 20 REPRESENTATIVE SAMPLES (UNOPENED) WERE RETURNED FOR INVESTIGATION. 10PCS FROM BATCH 8081043, 5PCS FROM 8116291 AND 5PCS 8060126. THE SAMPLES WERE SUBJECTED TO BLOOD ESCAPE TIME (BET) TEST. THE SAMPLES PASSED THE BET TEST, NO LEAKAGE WAS OBSERVED. A REVIEW OF PAST 12 MONTHS QUALITY NOTIFICATION WAS PERFORMED AND NO QUALITY NOTIFICATION WAS RAISED ON THE REPORTED DEFECT OF LEAKAGE. NO ABNORMALITY WAS DETECTED DURING THE PRODUCTION OF THE REPORTED BATCH. THE RETURNED SAMPLE PASSED THE BET TEST. HENCE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, THERE WAS A CAPA RAISED FOR BLOOD DROPLET FORMATION AT THE LUER END OF THE CATHETER ADAPTER (REFER TO CAPA#86125). SMALL ID CATHETER TUBING AND SMALL SEPTUM VENTS HAVE BEEN IMPLEMENTED TO ENHANCE THE BLOOD CONTROL FUNCTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD CATHENA¿ 22GX1.00IN STRAIGHT BC HAD INSTANCES OF VALVE NOT FUNCTIONING AFTER INSERTION OR CONNECTION/DISCONNECTION AS WELL AS LEAKAGE PAST THE VALVE AND OUT OF THE HUB ON 3 DIFFERENT BATCHES. INSERTION WAS OBSERVED ON LIVE PATIENT, STEPS FOR INSERTION WERE PERFORMED CORRECTLY, AND AS SOON AS THE NEEDLE WAS PULLED OUT OF THE HUB, THE BLOOD WOULD LEAK UP BY FOLLOWING THE SPRING INSIDE THE CATHETER HUB. CONNECTION WITH A SALINE FLUSH WAS PERFORMED, AFTER DISCONNECTION, CATHETER LEAKED = VALVE NOT FUNCTIONING EITHER. REPEATED THIS TWICE, WITH THE SAME RESULT.

Additional Manufacturer Narrative · 1

"MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8060126. MEDICAL DEVICE EXPIRATION DATE: 2021-02-28. DEVICE MANUFACTURE DATE: 2018-03-01. MEDICAL DEVICE LOT #: 8116291. MEDICAL DEVICE EXPIRATION DATE: 2021-04-30. DEVICE MANUFACTURE DATE: 2018-04-26. MEDICAL DEVICE LOT #: 8183043. MEDICAL DEVICE EXPIRATION DATE: 2021-06-30. DEVICE MANUFACTURE DATE: 2018-07-02. " (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT BD CATHENA¿ 22GX1.00IN STRAIGHT BC HAD INSTANCES OF VALVE NOT FUNCTIONING AFTER INSERTION OR CONNECTION/DISCONNECTION AS WELL AS LEAKAGE PAST THE VALVE AND OUT OF THE HUB ON 3 DIFFERENT BATCHES. INSERTION WAS OBSERVED ON LIVE PATIENT, STEPS FOR INSERTION WERE PERFORMED CORRECTLY, AND AS SOON AS THE NEEDLE WAS PULLED OUT OF THE HUB, THE BLOOD WOULD LEAK UP BY FOLLOWING THE SPRING INSIDE THE CATHETER HUB. CONNECTION WITH A SALINE FLUSH WAS PERFORMED, AFTER DISCONNECTION, CATHETER LEAKED = VALVE NOT FUNCTIONING EITHER. REPEATED THIS TWICE, WITH THE SAME RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186061 BD CATHENA¿ 22GX1.00IN STRAIGHT BC INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL (SINGAPORE) SEE H.10 00382903868063

Patients

Seq Age Sex Outcome Treatment
1 Other