BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER
Report
- Report Number
- 8041187-2019-00198
- Event Type
- Malfunction
- Date Received
- March 5, 2019
- Date of Event
- February 13, 2019
- Report Date
- March 27, 2019
- Manufacturer
- BECTON DICKINSON MEDICAL (SINGAPORE)
- Product Code
- FOZ
- UDI-DI
- 00382903868063
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: 20 REPRESENTATIVE SAMPLES (UNOPENED) WERE RETURNED FOR INVESTIGATION. 10PCS FROM BATCH 8081043, 5PCS FROM 8116291 AND 5PCS 8060126. THE SAMPLES WERE SUBJECTED TO BLOOD ESCAPE TIME (BET) TEST. THE SAMPLES PASSED THE BET TEST, NO LEAKAGE WAS OBSERVED. A REVIEW OF PAST 12 MONTHS QUALITY NOTIFICATION WAS PERFORMED AND NO QUALITY NOTIFICATION WAS RAISED ON THE REPORTED DEFECT OF LEAKAGE. NO ABNORMALITY WAS DETECTED DURING THE PRODUCTION OF THE REPORTED BATCH. THE RETURNED SAMPLE PASSED THE BET TEST. HENCE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, THERE WAS A CAPA RAISED FOR BLOOD DROPLET FORMATION AT THE LUER END OF THE CATHETER ADAPTER (REFER TO CAPA#86125). SMALL ID CATHETER TUBING AND SMALL SEPTUM VENTS HAVE BEEN IMPLEMENTED TO ENHANCE THE BLOOD CONTROL FUNCTION.
IT WAS REPORTED THAT ON A BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER THE, "ANTI-REFLUX VALVE DOESN'T FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION (MULTIGUARD NOT WORKING). THE REPORT IS AS FOLLOWS, "IT WAS REPORTED THE CATHETER ANTI REFLEX VALVE DOES NOT FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION AND THERE WAS LEAKING PAST THE VALVE AND OUT OF THE HUB. THIS COMPLAINT IS FOR PATIENT 8 OF 10. PER COMPLAINT FORM: CATHETER ANTI-REFLUX VALVE DOESN'T FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION (MULTIGUARD NOT WORKING). THERE IS BLOOD THAT IS LEAKING PAST THE VALVE AFTER REMOVAL OF THE NEEDLE, AND LEAKING OUT OF THE HUB. INSERTION WAS OBSERVED ON LIVE PATIENT, STEPS FOR INSERTION WERE PERFORMED CORRECTLY, AND AS SOON AS THE NEEDLE WAS PULLED OUT OF THE HUB, THE BLOOD WOULD LEAK UP BY FOLLOWING THE SPRING INSIDE THE CATHETER HUB. CONNECTION WITH A SALINE FLUSH WAS PERFORMED, AFTER DISCONNECTION, CATHETER LEAKED = VALVE NOT FUNCTIONING EITHER. REPEATED THIS TWICE, WITH THE SAME RESULT."
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8116291, MEDICAL DEVICE EXPIRATION DATE: 2021-04-30, DEVICE MANUFACTURE DATE: 2018-04-26. MEDICAL DEVICE LOT #: 8060126, MEDICAL DEVICE EXPIRATION DATE: 2021-02-28, DEVICE MANUFACTURE DATE: 2018-03-01. MEDICAL DEVICE LOT #: 8183043, MEDICAL DEVICE EXPIRATION DATE: 2021-06-30, DEVICE MANUFACTURE DATE: 2018-07-02. DEVICE EVALUATED BY MFR: A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT ON A BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER THE, ANTI-REFLUX VALVE DOESN'T FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION (MULTIGUARD NOT WORKING). THE REPORT IS AS FOLLOWS: "IT WAS REPORTED THE CATHETER ANTI REFLEX VALVE DOES NOT FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION AND THERE WAS LEAKING PAST THE VALVE AND OUT OF THE HUB. THIS COMPLAINT IS FOR PATIENT 8 OF 10. PER COMPLAINT FORM: CATHETER ANTI-REFLUX VALVE DOESN'T FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION (MULTIGUARD NOT WORKING). THERE IS BLOOD THAT IS LEAKING PAST THE VALVE AFTER REMOVAL OF THE NEEDLE, AND LEAKING OUT OF THE HUB. INSERTION WAS OBSERVED ON LIVE PATIENT, STEPS FOR INSERTION WERE PERFORMED CORRECTLY, AND AS SOON AS THE NEEDLE WAS PULLED OUT OF THE HUB, THE BLOOD WOULD LEAK UP BY FOLLOWING THE SPRING INSIDE THE CATHETER HUB. CONNECTION WITH A SALINE FLUSH WAS PERFORMED, AFTER DISCONNECTION, CATHETER LEAKED = VALVE NOT FUNCTIONING EITHER. REPEATED THIS TWICE, WITH THE SAME RESULT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186474 | BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER | INTRAVASCULAR CATHETER | FOZ | BECTON DICKINSON MEDICAL (SINGAPORE) | SEE H.10 | 00382903868063 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |