FDA Adverse Event Malfunction Summary report: N

BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY

MDR report key: 8391621 · Received March 5, 2019

Report

Report Number
8041187-2019-00193
Event Type
Malfunction
Date Received
March 5, 2019
Date of Event
February 13, 2019
Report Date
March 27, 2019
Manufacturer
BD (SUZHOU)
Product Code
FOZ
UDI-DI
00382903868063
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: 20 REPRESENTATIVE SAMPLES (UNOPENED) WERE RETURNED FOR INVESTIGATION. 10PCS FROM BATCH 8081043, 5PCS FROM 8116291 AND 5PCS 8060126. THE SAMPLES WERE SUBJECTED TO BLOOD ESCAPE TIME (BET) TEST. THE SAMPLES PASSED THE BET TEST, NO LEAKAGE WAS OBSERVED. A REVIEW OF PAST 12 MONTHS QUALITY NOTIFICATION WAS PERFORMED AND NO QUALITY NOTIFICATION WAS RAISED ON THE REPORTED DEFECT OF LEAKAGE. NO ABNORMALITY WAS DETECTED DURING THE PRODUCTION OF THE REPORTED BATCH. THE RETURNED SAMPLE PASSED THE BET TEST. HENCE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, THERE WAS A CAPA RAISED FOR BLOOD DROPLET FORMATION AT THE LUER END OF THE CATHETER ADAPTER (REFER TO CAPA#86125). SMALL ID CATHETER TUBING AND SMALL SEPTUM VENTS HAVE BEEN IMPLEMENTED TO ENHANCE THE BLOOD CONTROL FUNCTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY CATHETER ANTI REFLEX VALVE DOES NOT FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION AND THERE WAS LEAKING PAST THE VALVE AND OUT OF THE HUB.

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8116291. MEDICAL DEVICE EXPIRATION DATE: 2021-04-30. DEVICE MANUFACTURE DATE: 2018-04-26. MEDICAL DEVICE LOT #: 8060126. MEDICAL DEVICE EXPIRATION DATE: 2021-02-28. DEVICE MANUFACTURE DATE: 2018-03-01. MEDICAL DEVICE LOT #: 8183043. MEDICAL DEVICE EXPIRATION DATE: 2021-06-30. DEVICE MANUFACTURE DATE: 2018-07-02. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THE BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY CATHETER ANTI REFLEX VALVE DOES NOT FUNCTION AFTER INSERTION OR CONNECTION/DISCONNECTION AND THERE WAS LEAKING PAST THE VALVE AND OUT OF THE HUB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185220 BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY INTRAVASCULAR CATHETER FOZ BD (SUZHOU) SEE H.10 00382903868063

Patients

Seq Age Sex Outcome Treatment
1 Other