FDA Adverse Event Other Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 839084 · Received April 10, 2007

Report

Report Number
2245578-2007-00009
Event Type
Other
Date Received
April 10, 2007
Date of Event
March 12, 2007
Report Date
April 10, 2007
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

IN 2007, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT A PT WAS TREATED BASED ON THE FOLLOWING RESULTS FROM AN I-STAT CHEM8 CARTRIDGE LOT P06302; SODIUM 113 MEQ/L, POTASSIUM >9 MEQ/L, CHLORIDE 101 MEQ/L, TCO2 24 MEQ/L. PT WAS TREATED WITH INSULIN AND CALCIUM CHLORIDE, NA BICARB, D50, AND POSSIBLY KAYEXOLATE. A TUBE DRAWN AT THE SAME TIME (AS THE SAMPLE TESTED ON I-STAT THE SAME DAY) WAS SENT TO THE LAB AND YIELDED THE FOLLOWING RESULTS; SODIUM 123 MEQ/L, POTASSIUM 5.2 MEQ/L, CHLORIDE 88 MEQ/L, TCO2 24 MEQ/L. THERE WAS NO INJURY TO THE PT REPORTED. A POTASSIUM RESULT (LATER IN THE DAY) WAS 4.5 MEQ/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT CHEM8+ CARTRIDGE IVD MMI ABBOTT POINT OF CARE INC. NA P06302

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other