BD ULTRA-FINE¿ PEN NEEDLE

MDR report key: 8390688 · Received March 5, 2019

Report

Report Number
3006948883-2019-00156
Event Type
Malfunction
Date Received
March 5, 2019
Date of Event
February 11, 2019
Report Date
February 19, 2019
Manufacturer
BD (SUZHOU)
Product Code
FMI
PMA / PMN Number
K100005
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(6). INVESTIGATION SUMMARY: REFER TO (B)(4). AS PER MANUFACTURING, "DHR OF LOT 8123356 WAS REVIEWED AND NO QN FOUND. MANUFACTURE RECORDS WERE REVIEWED, NO ABNORMAL WAS FOUND. PEN NEEDLE PRODUCT HAS 100% CLOG TEST IN FLOWGURU STATION, AND DEFECT PART WILL REJECTED BY FLOWGURU STATION AUTOMATICALLY. CLOG TEST WAS CONDUCTED ON 10PCS RETENTION SAMPLES AND ALL NO NEEDLE CLOG FOUND. BASE ON INVESTIGATION ABOVE, THE COMPLAINT IS NOT MANUFACTURE ISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD ULTRA-FINE¿ PEN NEEDLE WOULD NOT DELIVER INSULIN. THIS OCCURRED ON 10 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED. VERBATIM: "I AM WRITING IN RELATION TO A BATCH OF BD ULTRAFINE 0.23 X 4MM NEEDLES. MY DAUGHTER WHO IS A (B)(6) YEAR OLD TYPE 1 DIABETIC USES THESE WITH A NOVORAPID PEN TO INJECT INSULIN. WE HAVE USED COUNTLESS NEEDLES AS SHE INJECTS 4 TIMES PER DAY AND WE HAVE JUST COME ACROSS OUR FIRST BATCH OF FAULTY NEEDLES. WHEN INSERTED, THE NEEDLES SEEM TO BE WORKING CORRECTLY, BUT BECAUSE OUR DAUGHTER HAS NOT FULLY PRIMED THE PEN WITH THE NEW NEEDLE, SHE PROCEEDS TO INJECT. THERE IS RESISTANCE IN THE PEN AND SHE REALISES THAT THE INSULIN IS NOT FLOWING FREELY. SOMETIMES SOME INSULIN COMES OUT, SOMETIMES NO INSULIN COMES OUT - IT IS DIFFICULT TO KNOW OR REGULATE. OBVIOUSLY THIS IS VERY DANGEROUS AS WE DO NOT KNOW HOW MUCH INSULIN SHE HAS TAKEN AND WE HAVE TO BE EXTREMELY CAREFUL NOT TO OVERDOSE INSULIN. WHEN THESE ARE FIRST INSERTED, SOME INSULIN WILL COLLECT AND SPILL FROM THE POINT OF THE NEEDLE, LEADING ONE TO BELIEVE THAT IT IS WORKING CORRECTLY BUT THEN THE NEEDLE WILL NOT ALLOW MORE INSULIN THROUGH. THERE HAVE BEEN APPROXIMATELY 20 NEEDLES THAT HAVE BEEN FAULTY. OF THE LAST FEW THAT HAVE HAD ISSUES, I HAVE KEPT THE SERIAL NUMBERS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
185798 BD ULTRA-FINE¿ PEN NEEDLE INSULIN PEN NEEDLE FMI BD (SUZHOU) 8123356

Patients

Seq Age Sex Outcome Treatment
1 Other