COMPR SRS ANTI ROT IC SEG-30MM
Report
- Report Number
- 0001825034-2019-00985
- Event Type
- Injury
- Date Received
- March 5, 2019
- Date of Event
- December 17, 2018
- Report Date
- June 27, 2019
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KWT
- PMA / PMN Number
- K111746
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE FOLLOWING REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION RECEIVED: THE COMPLAINT WAS CONFIRMED BASED ON THE OPERATIVE NOTES THAT WAS PROVIDED STATING DRAINING QUITE A BIT FROM OPENING AT THE APEX OF THE INCISION WHERE THE DISTAL HUMERAL COMPONENT IS NEARLY SUBCU. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO RELATED DEVIATIONS OR ANOMALIES. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO ADDITIONAL INFORMATION IS REPORTED AT THIS TIME.
(B)(4). CONCOMITANT MEDICAL PRODUCTS: 110029939 COMPR SRS 60MM DST HUM BDY RT 588380; 211269 COMPR SRS SMALL FLANGE 155830; 211236 COMPR SRS MOD STEM - 8X100MM 365290; 00840002411 ULNAR COMPONENT PLASMA SPRAYED SIZE 4 115 MM LENGTH RIGHT FOR CEMENTED USE ONLY 63269429. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, DISCARDED. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 -2019 -00984; 0001825034 -2019 -00987; 0001825034 -2019 -00988.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE TO REMOVE BROKEN IMPLANT AND WAS SUBSEQUENTLY REVISED TO REMOVE THE REMAINING HARDWARE DUE TO INFECTION APPROXIMATELY 3 WEEKS POST REVISION SURGERY. DRAINAGE OF THE SURGICAL SITE AND VISIBLE, EXPOSED HARDWARE WAS NOTED THROUGH THE INCISION SITE. NO ADDITIONAL INFORMATION IS AVAILABLE FROM THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 184077 | COMPR SRS ANTI ROT IC SEG-30MM | PROSTHESIS, SHOULDER | KWT | ZIMMER BIOMET, INC. | N/A | 351820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |