ARCHITECT I2000SR ANALYZER
Report
- Report Number
- 1628664-2019-00176
- Event Type
- Malfunction
- Date Received
- March 2, 2019
- Date of Event
- February 13, 2019
- Report Date
- March 2, 2019
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- UDI-DI
- 00380740006235
- PMA / PMN Number
- K983212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
WHILE TROUBLESHOOTING THE ISSUE, THE FIELD SERVICE ENGINEER DISCOVERED DAMAGE (CHARRING) OF THE CONTROLLER PWB ON A SMALL AREA AROUND A CAPACITOR ON THE BOARD. THE PART WAS REPLACED TO RESOLVE THE ISSUE. A REVIEW OF THE ANALYZER SERVICE HISTORY FOUND NO ISSUES THAT MAY HAVE CONTRIBUTED TO THE EVENT. THERE WERE NO SUBSEQUENT REPORTS OF SMOKING/BURN ISSUES SINCE THE PART WAS REPLACED. THE 2019 UL CERTIFICATION MEMO INDICATES THAT ABBOTT DIAGNOSTIC EQUIPMENT AND ACCESSORIES ARE CERTIFIED TO THE APPROPRIATE SAFETY STANDARDS, AND ADEQUATE PROTECTION IS PROVIDED FOR THE OPERATOR AGAINST SPREAD OF FIRE FROM THE EQUIPMENT. THE BURN DAMAGE, OBSERVED ON THE CONTROLLER PWB WAS LIMITED TO A SMALL AREA AROUND CAPACITOR C10 ON THE BOARD NEAR FUSE 2; THE DAMAGE DID NOT SPREAD TO OTHER PARTS OF THE INSTRUMENT. A REVIEW OF COMPLAINT AND TRENDING DATA DID NOT IDENTIFY ANY ADVERSE TRENDS FOR TICKETS WITH REPLACEMENTS OF THE CONTROLLER PWB. DURING A 91 MONTH SEARCH FOR SIMILAR COMPLAINTS, NO OTHER COMPLAINTS WERE IDENTIFIED RELATED TO CHARRING ON THE CONTROLLER PWB. A REVIEW OF ARCHITECT I2000SR TRACKING AND TRENDING DATA IN THE PRODUCT MONITORING REVIEW REVEALED NO SYSTEMIC ISSUES OR ADVERSE TRENDS RELATED TO THE ISSUE IDENTIFIED IN THIS COMPLAINT. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER'S ISSUE. NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT I2000SR WAS IDENTIFIED.
THE ABBOTT FIELD SERVICE REPRESENTATIVE REPORTED WHILE REPLACING THE IARM CONTROLLER PCW BOARD ON THE ARCHITECT I2000SR ANALYZER, SMOKE WAS OBSERVED. THERE WERE NO INJURIES AND NO IMPACT TO PATIENT MANAGEMENT OR USER SAFETY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 177504 | ARCHITECT I2000SR ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | JJE | ABBOTT MANUFACTURING INC | 00380740006235 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CONTROLLER PWB , PART 7-900505-01 |