FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY

MDR report key: 8383816 · Received March 1, 2019

Report

Report Number
3013756811-2019-10716
Event Type
Malfunction
Date Received
March 1, 2019
Date of Event
February 8, 2019
Report Date
March 1, 2019
Manufacturer
TANDEM DIABETES CARE
Product Code
OZO
UDI-DI
00853052007264
PMA / PMN Number
P180008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. ADDITIONALLY, THE BATTERY GAUGE WAS OBSERVED TO BE FLUCTUATING. THE CUSTOMER'S BLOOD GLUCOSE (BG) RANGED FROM 190-281 MG/DL. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER WAS INSTRUCTED TO LEAVE THE PUMP PLUGGED INTO A POWER SOURCE FOR AT LEAST 30 MINUTES. CUSTOMER ACKNOWLEDGED. UPON FOLLOW UP, THE CUSTOMER REPORTED THAT THE ALERT CLEARED AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
175845 T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO TANDEM DIABETES CARE 1000096 00853052007264

Patients

Seq Age Sex Outcome Treatment
1 51 YR