FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 8380655 · Received February 28, 2019

Report

Report Number
2531779-2019-01616
Event Type
Malfunction
Date Received
February 28, 2019
Report Date
February 25, 2019
Manufacturer
ANIMAS LLC
Product Code
OYC
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 25-FEB-2019. WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED CONTAMINATION UNDER ALL OF THE KEYPAD BUTTONS. THE KEYPAD COVER WAS FOUND TO BE PEELING FROM THE BASE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED CONTAMINATION UNDER ALL OF THE KEYPAD BUTTONS. THE KEYPAD COVER WAS FOUND TO BE PEELING FROM THE BASE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 25-FEB-2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
170414 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS LLC

Patients

Seq Age Sex Outcome Treatment
1 49 YR