FDA Adverse Event Other Summary report: N

OPEN HEART PACK

MDR report key: 83805 · Received April 15, 1997

Report

Report Number
1423537-1997-00062
Event Type
Other
Date Received
April 15, 1997
Date of Event
March 20, 1997
Report Date
April 15, 1997
Manufacturer
ALLEGIANCE-SURGICAL GROUP
Product Code
GDY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALLEGIANCE RESPONSE: BATCH HISTORY RECORD REVIEWED FOR ID OF PRODUCTS INVOLVED. NO DEVIATIONS WERE NOTED DURING THE MFG PROCESS. NO ADD'L HANDLING OF THESE COMPONENTS IS DONE DURING ASSEMBLY. SUPPLIERS NOTIFIED. VALLEYLAB INC RESPONSE: A SEARCH OF THE PRODUCT HISTORY WAS PERFORMED FOR THE LOT NUMBER REPORTED AND NO PERTINENT ENTRIES WERE FOUND. NO PRODUCT WAS RETURNED FOR EVAL, THEREFORE NO TESTING COULD BE PERFORMED. JOHNSON & JOHNSON MEDICAL INC (FOR MEDICAL TASK FORCE) RESPONSE: WITHOUT A SAMPLE, A COMPLETE INVESTIGATION IS NOT POSSIBLE. HOWEVER, THIS SPONGE IS COMPOSED OF COTTON AND UNDER CERTAIN CIRCUMSTANCES (HEAT SOURCE AND OXYGEN) WILL IGNITE. COMPLAINTS ARE REVIEWED BY THE MFG SITE AND TRENDS ARE MONITORED.

Description of Event or Problem · 1

THE PHYSICIAN WAS TAKING DOWN THE MAMMARY ARTERY, USING A 4X4 X-RAY SPONGE AND CAUTERY PENCIL, WHEN THE SPONGE CAUGHT FIRE. THE SPONGE WAS THROWN ON THE FLOOR AND AN ASEPTO OF WATER WAS USED TO PUT OUT THE FIRE. THERE WERE NO INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPEN HEART PACK OPEN HEART PACK GDY ALLEGIANCE-SURGICAL GROUP SCV38OHSW 222057

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other