COMPLETE WATER PURIFICATION SYSTEM
Report
- Report Number
- 3019131-2019-00003
- Event Type
- Injury
- Date Received
- February 28, 2019
- Date of Event
- January 29, 2019
- Report Date
- February 28, 2019
- Manufacturer
- MAR COR PURIFICATION
- Product Code
- FIP
- PMA / PMN Number
- K945559
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- 003
Narratives
THE FACILITY REPORTED A PRIVATE PLUMBER INSTALLED A BRASS FITTING IN THE WATER LOOP OF THEIR COMPLETE WATER PURIFICATION SYSTEM WHEN SPLICING THE LOOP TOGETHER AFTER RELOCATING IT. THERE IS POTENTIAL FOR BRASS TO LEACH DOWNSTREAM THROUGH THE REVERSE OSMOSIS (RO) PATHWAY, THUS POTENTIAL FOR PATIENT EXPOSURE DURING HEMODIALYSIS TREATMENTS. IT WAS REPORTED THAT THE FACILITY RAN ONE DAY OF PATIENT TREATMENTS AFTER THE BRASS FITTING WAS INSTALLED BY THE PLUMBER. AFTER THE FACILITY'S BIOMED NOTICED THE BRASS FITTING, THEY REPORTED REPLACING THE BRASS FITTING WITH A PVC OR PEX FITTING. ANSI/AAMI 26722:2014 'WATER TREATMENT EQUIPMENT FOR HEMODIALYSIS AND RELATED THERAPIES' SPECIFICALLY STATES THAT BRASS MATERIALS SHOULD BE AVOIDED DUE TO BEING DOCUMENTED AS BEING HAZARDOUS TO THE PATIENT. ADDITIONALLY, IF CHANGES ARE MADE AFTER INSTALLATION OF THE SYSTEM, THE FACILITY IS RESPONSIBLE FOR ENSURING APPROPRIATE MATERIALS/COMPONENTS ARE USED. WHILE THE RISK TO PATIENTS DERIVES FROM LONG-TERM CUMULATIVE TOXICITY, MAR COR PURIFICATION HAS DECIDED TO CONSERVATIVELY SUBMIT THIS REPORT DUE TO THE POTENTIAL FOR PATIENT EXPOSURE TO BRASS LEACHING INTO THE RO PATHWAY. THERE HAVE BEEN NO REPORTS OF PATIENT HARM. THIS COMPLAINT WILL CONTINUE BEING MONITORED IN THE MAR COR PURIFICATION COMPLAINT HANDLING SYSTEM.
THE FACILITY REPORTED A PRIVATE PLUMBER INSTALLED A BRASS FITTING IN THE WATER LOOP OF THEIR COMPLETE WATER PURIFICATION SYSTEM WHEN SPLICING THE LOOP TOGETHER AFTER RELOCATING IT. THERE IS POTENTIAL FOR BRASS TO LEACH DOWNSTREAM THROUGH THE REVERSE OSMOSIS (RO) PATHWAY, THUS POTENTIAL FOR PATIENT EXPOSURE DURING HEMODIALYSIS TREATMENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173423 | COMPLETE WATER PURIFICATION SYSTEM | WATER PURIFICATION SYSTEM FOR HEMODIALYSIS | FIP | MAR COR PURIFICATION | PVC LOOP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |