FDA Adverse Event Injury Summary report: N

COMPLETE WATER PURIFICATION SYSTEM

MDR report key: 8380256 · Received February 28, 2019

Report

Report Number
3019131-2019-00003
Event Type
Injury
Date Received
February 28, 2019
Date of Event
January 29, 2019
Report Date
February 28, 2019
Manufacturer
MAR COR PURIFICATION
Product Code
FIP
PMA / PMN Number
K945559
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY REPORTED A PRIVATE PLUMBER INSTALLED A BRASS FITTING IN THE WATER LOOP OF THEIR COMPLETE WATER PURIFICATION SYSTEM WHEN SPLICING THE LOOP TOGETHER AFTER RELOCATING IT. THERE IS POTENTIAL FOR BRASS TO LEACH DOWNSTREAM THROUGH THE REVERSE OSMOSIS (RO) PATHWAY, THUS POTENTIAL FOR PATIENT EXPOSURE DURING HEMODIALYSIS TREATMENTS. IT WAS REPORTED THAT THE FACILITY RAN ONE DAY OF PATIENT TREATMENTS AFTER THE BRASS FITTING WAS INSTALLED BY THE PLUMBER. AFTER THE FACILITY'S BIOMED NOTICED THE BRASS FITTING, THEY REPORTED REPLACING THE BRASS FITTING WITH A PVC OR PEX FITTING. ANSI/AAMI 26722:2014 'WATER TREATMENT EQUIPMENT FOR HEMODIALYSIS AND RELATED THERAPIES' SPECIFICALLY STATES THAT BRASS MATERIALS SHOULD BE AVOIDED DUE TO BEING DOCUMENTED AS BEING HAZARDOUS TO THE PATIENT. ADDITIONALLY, IF CHANGES ARE MADE AFTER INSTALLATION OF THE SYSTEM, THE FACILITY IS RESPONSIBLE FOR ENSURING APPROPRIATE MATERIALS/COMPONENTS ARE USED. WHILE THE RISK TO PATIENTS DERIVES FROM LONG-TERM CUMULATIVE TOXICITY, MAR COR PURIFICATION HAS DECIDED TO CONSERVATIVELY SUBMIT THIS REPORT DUE TO THE POTENTIAL FOR PATIENT EXPOSURE TO BRASS LEACHING INTO THE RO PATHWAY. THERE HAVE BEEN NO REPORTS OF PATIENT HARM. THIS COMPLAINT WILL CONTINUE BEING MONITORED IN THE MAR COR PURIFICATION COMPLAINT HANDLING SYSTEM.

Description of Event or Problem · 1

THE FACILITY REPORTED A PRIVATE PLUMBER INSTALLED A BRASS FITTING IN THE WATER LOOP OF THEIR COMPLETE WATER PURIFICATION SYSTEM WHEN SPLICING THE LOOP TOGETHER AFTER RELOCATING IT. THERE IS POTENTIAL FOR BRASS TO LEACH DOWNSTREAM THROUGH THE REVERSE OSMOSIS (RO) PATHWAY, THUS POTENTIAL FOR PATIENT EXPOSURE DURING HEMODIALYSIS TREATMENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173423 COMPLETE WATER PURIFICATION SYSTEM WATER PURIFICATION SYSTEM FOR HEMODIALYSIS FIP MAR COR PURIFICATION PVC LOOP

Patients

Seq Age Sex Outcome Treatment
1 Other