FDA Adverse Event Injury Summary report: N

STEM AMISTEM H CEMENTLESS SIZE 7

MDR report key: 8377331 · Received February 28, 2019

Report

Report Number
3005180920-2019-00101
Event Type
Injury
Date Received
February 28, 2019
Date of Event
January 29, 2019
Report Date
February 28, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804137
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON (B)(6) 2019: LOT 131576: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-JUL-2013. EXPIRATION DATE: 2018-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: HIP REVISION SURGERY OCCURRED 5 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD MAN. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF OSTEOLYTIC REGIONS IN GRUEN ZONES 1 AND 7 AND SIGNS OF STRESS-SHIELDING. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS TOTAL HIP REPLACEMENTS AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS EVENT CANNOT BE DETERMINED.

Description of Event or Problem · 1

ON (B)(6) 2019 WE WERE INFORMED ABOUT A REVISION SURGERY, PLANNED FOR (B)(6) 2019, SO 5 YEARS AND TWO MONTHS AFTER PRIMARY SURGERY, DUE TO STEM LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173061 STEM AMISTEM H CEMENTLESS SIZE 7 HIP STEM CEMENTLESS LZO MEDACTA INTERNATIONAL SA 131576 07630030804137

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention