STEM AMISTEM H CEMENTLESS SIZE 7
Report
- Report Number
- 3005180920-2019-00101
- Event Type
- Injury
- Date Received
- February 28, 2019
- Date of Event
- January 29, 2019
- Report Date
- February 28, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030804137
- PMA / PMN Number
- K093944
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON (B)(6) 2019: LOT 131576: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-JUL-2013. EXPIRATION DATE: 2018-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: HIP REVISION SURGERY OCCURRED 5 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN A (B)(6) YEAR OLD MAN. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF OSTEOLYTIC REGIONS IN GRUEN ZONES 1 AND 7 AND SIGNS OF STRESS-SHIELDING. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS TOTAL HIP REPLACEMENTS AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS EVENT CANNOT BE DETERMINED.
ON (B)(6) 2019 WE WERE INFORMED ABOUT A REVISION SURGERY, PLANNED FOR (B)(6) 2019, SO 5 YEARS AND TWO MONTHS AFTER PRIMARY SURGERY, DUE TO STEM LOOSENING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173061 | STEM AMISTEM H CEMENTLESS SIZE 7 | HIP STEM CEMENTLESS | LZO | MEDACTA INTERNATIONAL SA | 131576 | 07630030804137 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention |