FDA Adverse Event Malfunction Summary report: N

OT VERIO FLEX METER

MDR report key: 8376998 · Received February 28, 2019

Report

Report Number
3008382007-2019-00559
Event Type
Malfunction
Date Received
February 28, 2019
Report Date
February 25, 2019
Manufacturer
LIFESCAN EUROPE GMBH
Product Code
NBW
UDI-DI
00353885010986
PMA / PMN Number
K150214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Description of Event or Problem · 1

ON (B)(6) 2019, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE ERRATIC RESULTS ON THE SUBJECT METER OF "90, 147, 140, 125 AND 89 MG/DL", PERFORMED WITHIN 20 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172622 OT VERIO FLEX METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE GMBH 4482778 00353885010986

Patients

Seq Age Sex Outcome Treatment
1