FDA Adverse Event Malfunction Summary report: N

DEFENSE BLOOD STANDARD SYSTEM

MDR report key: 837594 · Received February 23, 2007

Report

Report Number
1119523-2007-00001
Event Type
Malfunction
Date Received
February 23, 2007
Date of Event
December 21, 2006
Report Date
February 22, 2007
Manufacturer
DOD/DBSS/PROJECT OFFICE
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DBSS, UNDER CERTAIN CONDITIONS DOES NOT CALCULATE THE EXPIRATION DATE OF CONVERTED PRODUCTS CORRECTLY. NO FACILITY HAS REPORTED ISSUING OR TRANSFUSING ANY EXPIRED PRODUCTS DUE TO THE SOFTWARE ANOMALY. DBSS TECHNICAL BULLETIN (DTB) 06-12 WAS ISSUED TO ALERT USERS OF TWO KNOWN SCENARIOS WHERE DBSS WILL INCORRECTLY CALCULATE THE EXPIRATION DATE WHEN CONVERTING PRODUCTS. ATTACHMENT: DTB 06-12.

Description of Event or Problem · 1

DBSS EXTENDED THE EXPIRATION DATE OF AN IRRADIATED PLATELET PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEFENSE BLOOD STANDARD SYSTEM BLOOD ESTABLISHMENT COMPUTER SOFTWARE MMH DOD/DBSS/PROJECT OFFICE * *

Patients

Seq Age Sex Outcome Treatment
1 *