FDA Adverse Event Injury Summary report: N

ASR TAP SLV ADAP 12/14 -1

MDR report key: 8375303 · Received February 27, 2019

Report

Report Number
1818910-2019-85842
Event Type
Injury
Date Received
February 27, 2019
Date of Event
February 4, 2019
Report Date
February 4, 2019
Manufacturer
DEPUY INTERNATIONAL LTD. 801037
Product Code
KXA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISION OF 21 CFR, PART 803. THE REPORT MAY BE BASED ON THE INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: CORRECTED: H6(DEVICE CODE). PRODUCT COMPLAINT (B)(4). INVESTIGATION SUMMARY : NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THIS HIP REPLACEMENT PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET AND THE PRODUCT CODES ARE NOW CONSIDERED INACTIVE. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PATIENT CODE: NO CODE AVAILABLE (3191) USED TO CAPTURE THE MEDICAL DEVICE REMOVAL.

Description of Event or Problem · 0

AFTER REVIEW OF MEDICAL RECORDS THE PATIENT WAS REVISED TO ADDRESS PAIN AND ELEVATED METAL IONS. OPERATIVE FINDING REPORTED HETEROTOPIC OSSIFICATION, METALLOSIS IN THE CAPSULE, STAINING OF THE TRUNNION FEMORAL COMPONENT WITH PROXIMAL OSTEOLYSIS AND NO BONY INGROWTH IN THE ACETABULAR COMPONENT. SURGICAL PROLONG FOR 45-50 MINUTES DUE TO OBESITY AND EXTENSIVE HO.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY ==> THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN. THE INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS WAS CONDUCTED UNDER MDD CAPA-(B)(4). ONGOING POST MARKET SURVEILLANCE IS CONDUCTED PER OUR PROCEDURES FOR THIS PRODUCT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT #: (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Description of Event or Problem · 1

LITIGATION ALLEGED THE PATIENT SUFFERED PAIN, PERMANENT PHYSICAL INJURIES, DISFIGUREMENT AND EXPOSURE TO EXCESSIVE LEVELS OF CHROMIUM AND COBALT AS A RESULT OF THE IMPLANTED ASR HIP. UPDATE AD 04 FEB 2019: RECEIPT OF DER. THERE WERE NO NEW ALLEGATION STATED HOWEVER, PATIENT HAS BEEN REVISED DUE TO FAILED OF THE ASR XL ACETABULAR IMPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167091 ASR TAP SLV ADAP 12/14 -1 ASR HIP SYSTEM : HIP FEMORAL SLEEVE KXA DEPUY INTERNATIONAL LTD. 801037

Patients

Seq Age Sex Outcome Treatment
1 82 YR Male Required Intervention ASR ACETABULAR CUPS 52.| ASR UNI FEMORAL IMPL SIZE 46.| CORAIL2 STD SIZE 14.