FDA Adverse Event Malfunction Summary report: N

5.0MM CANNULATED VA LOCKING SCREW/80MM

MDR report key: 8374731 · Received February 27, 2019

Report

Report Number
2939274-2019-56699
Event Type
Malfunction
Date Received
February 27, 2019
Date of Event
February 2, 2019
Report Date
February 2, 2019
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HWC
UDI-DI
10886982062918
PMA / PMN Number
K110354
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL PRODUCT CODES: JDP AND HRS. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6) 2019, THE SURGEON NOTED THAT THE CANNULATED VARIABLE LOCKING SCREW DID NOT LOCK INTO DISTAL HOLE AND THE SURGEON ALSO MENTIONED IT MAY HAVE STRIPPED THE THREADS ON THE VARIABLE ANGLE- LOCKING CONDYLAR PLATE DURING SCREW INSERTION. FRAGMENTS WERE REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THERE WAS NO SURGICAL DELAY AND NO PATIENT CONSEQUENCE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR A CANNULATED VARIABLE LOCKING SCREW. THIS IS REPORT 1 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167284 5.0MM CANNULATED VA LOCKING SCREW/80MM SCREW, FIXATION, BONE HWC WRIGHTS LANE SYNTHES USA PRODUCTS LLC 02.231.680 10886982062918

Patients

Seq Age Sex Outcome Treatment
1 59 YR