NEUROFORM MICRODELIVERY STENT SYSTEM
Report
- Report Number
- 6000078-2007-00055
- Event Type
- Malfunction
- Date Received
- February 16, 2007
- Date of Event
- January 23, 2007
- Report Date
- January 23, 2007
- Manufacturer
- NEUROVASCULAR, A DIVISION OF BOSTON SCIENTIFIC CORP.
- Product Code
- NJE
- PMA / PMN Number
- H020002/S4
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL PMA/510 (K): H020002/S5.
IT WAS REPORTED THAT THE PHYSICIAN WAS ATTEMPTING TO PLACE A STENT ACROSS THE NECK OF A GIANT CAVERNOUS INTERNAL CAROTID ARTERY (ICA) ANEURYSM FOR COILING. THE STENT SYSTEM WAS ADVANCED ACROSS THE NECK, BUT THE STENT COULD NOT BE DELIVERED. THE STENT HAD BEEN ADVANCED WITH THE USE OF A 300CM FLOPPY GUIDEWIRE. WHEN THE ATTEMPT WAS MADE TO DRAW BACK THE DELIVERY CATHETER, IT WAS DISCOVERED THAT THE STENT WAS NOT BEING UNSHEATHED. IT WAS FELT THAT THE STABILIZER WAS NOT LONG ENOUGH TO REACH THE END OF THE DELIVERY SYSTEM, AND THAT A FAIR AMOUNT OF FRICTION HAD BEEN FELT. UPON REMOVAL OF THE SYSTEM, IT WAS SEEN THAT THE 3 FRENCH DELIVERY CATHETER HAD BROKEN ABOUT 10 CMS FROM THE HUB ALONG THE CATHETER, AND THUS WOULD NOT DRAW BACK IN ORDER TO DELIVER THE STENT. THE ENTIRE STENT SYSTEM WAS DISCARDED. IT WAS REPORTED THE WIRE HAVE BEEN KINKED IN HANDLING, WHICH MAY HAVE CAUSED SOME FRICTION. A THIRD SYSTEM WAS PREPPED, POSITIONED ALONG A .012 WIRE, AND AGAIN THE DELIVERY CATHETER STRETCHED. THE STENT WAS WELL POSITIONED, SO THEY WERE ABLE TO PLACE THE STENT BY USING A . 018 WIRE AS THE STABILIZER. THE ANEURYSM WAS COILED TO THE PHYSICIAN'S SATISFACTION AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. THE PATIENT WAS REPORTED TO BE FINE FOLLOWING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEUROFORM MICRODELIVERY STENT SYSTEM | (NJE) INTRACRANIAL STENT | NJE | NEUROVASCULAR, A DIVISION OF BOSTON SCIENTIFIC CORP. | SNF34030 | 8517408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | GDC 360 COILS| TRANSCEND FLOPPY GUIDEWIRE 300CM| THRUWAY WIRE .018| HEADLINER WIRE .012| 3 FRENCH DELIVERY CATHETER (MODEL/LOT NUMBER UNK) |