FDA Adverse Event Malfunction Summary report: N

NEUROFORM MICRODELIVERY STENT SYSTEM

MDR report key: 837384 · Received February 16, 2007

Report

Report Number
6000078-2007-00055
Event Type
Malfunction
Date Received
February 16, 2007
Date of Event
January 23, 2007
Report Date
January 23, 2007
Manufacturer
NEUROVASCULAR, A DIVISION OF BOSTON SCIENTIFIC CORP.
Product Code
NJE
PMA / PMN Number
H020002/S4
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL PMA/510 (K): H020002/S5.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN WAS ATTEMPTING TO PLACE A STENT ACROSS THE NECK OF A GIANT CAVERNOUS INTERNAL CAROTID ARTERY (ICA) ANEURYSM FOR COILING. THE STENT SYSTEM WAS ADVANCED ACROSS THE NECK, BUT THE STENT COULD NOT BE DELIVERED. THE STENT HAD BEEN ADVANCED WITH THE USE OF A 300CM FLOPPY GUIDEWIRE. WHEN THE ATTEMPT WAS MADE TO DRAW BACK THE DELIVERY CATHETER, IT WAS DISCOVERED THAT THE STENT WAS NOT BEING UNSHEATHED. IT WAS FELT THAT THE STABILIZER WAS NOT LONG ENOUGH TO REACH THE END OF THE DELIVERY SYSTEM, AND THAT A FAIR AMOUNT OF FRICTION HAD BEEN FELT. UPON REMOVAL OF THE SYSTEM, IT WAS SEEN THAT THE 3 FRENCH DELIVERY CATHETER HAD BROKEN ABOUT 10 CMS FROM THE HUB ALONG THE CATHETER, AND THUS WOULD NOT DRAW BACK IN ORDER TO DELIVER THE STENT. THE ENTIRE STENT SYSTEM WAS DISCARDED. IT WAS REPORTED THE WIRE HAVE BEEN KINKED IN HANDLING, WHICH MAY HAVE CAUSED SOME FRICTION. A THIRD SYSTEM WAS PREPPED, POSITIONED ALONG A .012 WIRE, AND AGAIN THE DELIVERY CATHETER STRETCHED. THE STENT WAS WELL POSITIONED, SO THEY WERE ABLE TO PLACE THE STENT BY USING A . 018 WIRE AS THE STABILIZER. THE ANEURYSM WAS COILED TO THE PHYSICIAN'S SATISFACTION AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. THE PATIENT WAS REPORTED TO BE FINE FOLLOWING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEUROFORM MICRODELIVERY STENT SYSTEM (NJE) INTRACRANIAL STENT NJE NEUROVASCULAR, A DIVISION OF BOSTON SCIENTIFIC CORP. SNF34030 8517408

Patients

Seq Age Sex Outcome Treatment
1 67 YR GDC 360 COILS| TRANSCEND FLOPPY GUIDEWIRE 300CM| THRUWAY WIRE .018| HEADLINER WIRE .012| 3 FRENCH DELIVERY CATHETER (MODEL/LOT NUMBER UNK)