FDA Adverse Event
Injury
Summary report: N
*
MDR report key: 837228
·
Received April 12, 2007
Report
- Report Number
- 837228
- Event Type
- Injury
- Date Received
- April 12, 2007
- Date of Event
- January 29, 2005
- Report Date
- April 12, 2007
- Manufacturer
- CHIU TECHNICAL CORP
- Product Code
- HJM
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- DC, US
Narratives
Description of Event or Problem · 1
NEWBORN RECENTLY TRANSFERRED TO NICU FROM THE BIRTH HOSPITAL, WAS FOUND TO HAVE BLISTERING ALONG THE SIDE OF THE RIGHT LEG AFTER THE DRAPES WERE REMOVED FOLLOWING PICC LINE PLACEMENT. A TRANSILLUMINATOR HAD BEEN USED TO FACILITATE VEIN LOCATION. THE DEVICE WAS TESTED BY OUR BIOMED DEPT AND IT APPEARED TO BE WORKING CORRECTLY BUT IT WAS NOTED THAT THE BOX PART GOT VERY HOT IF COVERED BY A BLANKET OR TOWEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | * | LIGHT SOURCE | HJM | CHIU TECHNICAL CORP | BFO-150 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 DA | Required Intervention |