BD VACUTAINER® K2E (EDTA) 7.2MG PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2019-00236
- Event Type
- Malfunction
- Date Received
- February 26, 2019
- Date of Event
- February 8, 2019
- Report Date
- April 1, 2019
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 50382903688611
- PMA / PMN Number
- BK050036
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT PHOTOS WERE PROVIDED BY THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. ROOT CAUSE DESCRIPTION: A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. CAPA 161123. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.
IT WAS REPORTED THAT BD VACUTAINER® K2E (EDTA) 7.2MG PLUS BLOOD COLLECTION TUBES HAD UNDERFILL. THIS OCCURRED ON 200 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8040575. MEDICAL DEVICE EXPIRATION DATE: 2019-06-30. DEVICE MANUFACTURE DATE: 2018-02-09. MEDICAL DEVICE LOT #: 8099519. MEDICAL DEVICE EXPIRATION DATE: 2019-07-31. DEVICE MANUFACTURE DATE: 2018-04-09. (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD VACUTAINER® K2E (EDTA) 7.2MG PLUS BLOOD COLLECTION TUBES HAD UNDERFILL. THIS OCCURRED ON 200 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 162818 | BD VACUTAINER® K2E (EDTA) 7.2MG PLUS BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | SEE H.10 | 50382903688611 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |