EAGLE EYE GOLD CATHETER
MDR report key: 836945
·
Received April 10, 2007
Report
- Report Number
- 2939520-2007-00005
- Event Type
- Injury
- Date Received
- April 10, 2007
- Date of Event
- February 20, 2007
- Report Date
- April 9, 2007
- Manufacturer
- VOLCANO CORPORATION
- Product Code
- ITX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE PHYSICIAN USED THE EAGLE GOLD CATHETER TO EXAMINE THE SHUNTING POINT OF A DIALYSIS PATIENT. THE RADIAL ARTERY WAS PUNCTURED AND A 6F SHEATH WAS INSERTED INTO THE ARTERY. THE CATHETER BECAME TRAPPED AT THE TIP OF THE SHEATH AND DETACHED. A SNARED WAS ATTEMPTED, BUT UNSUCCESSFUL. THE PATIENT WAS TAKEN TO SURGERY AND THE DETACHED TIP WAS REMOVED. PATIENT WAS REPORTED TO BE IN GOOD CONDITION. THE CATHETER HAS NOT RETURNED FOR EVALUATION; THEREFORE, FURTHER ANALYSIS CANNOT BE COMPLETED AT THIS TIME.
Description of Event or Problem · 1
DISTAL TIP OF CATHETER TRAPPED AT THE TIP OF THE SHEATH AND DETACHED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EAGLE EYE GOLD CATHETER | INTRAVASCULAR ULTRASONIC CATHETER | ITX | VOLCANO CORPORATION | 85900 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Required Intervention |